张辉1, 2 王亚东1, 2 马彦1, 等. 塑料闪烁微球的背景计数率和计数效率研究[J/OL]. 核技术, 2025,48.
Zhang Hui1, 2 Wang Yadong1, 2 Ma Yan1, et al. Study on Background Count Rate and Counting Efficiency of Plastic Scintillation Microspheres[J/OL]. Nuclear techniques, 2025, 48.
张辉1, 2 王亚东1, 2 马彦1, 等. 塑料闪烁微球的背景计数率和计数效率研究[J/OL]. 核技术, 2025,48. DOI: 10.11889/j.0253-3219.2025.hjs.48.240277.
Zhang Hui1, 2 Wang Yadong1, 2 Ma Yan1, et al. Study on Background Count Rate and Counting Efficiency of Plastic Scintillation Microspheres[J/OL]. Nuclear techniques, 2025, 48. DOI: 10.11889/j.0253-3219.2025.hjs.48.240277.
[Background]: Plastic scintillation microspheres (PSm) are a novel luminescent material with significant potential in the measurement and continuous monitoring of radionuclide activity. [Purpose]: To optimize the application of PSm in radionuclide activity analysis
it is crucial to understand the impact of different measurement conditions on the background count rate and counting efficiency of PSm. [Methods]: The chemiluminescence and photoluminescence of PSm were measured using a Hidex 300SL liquid scintillation counter. The study also investigated the effect of PSm dosage and the type of liquid scintillation vial on the background count rate of PSm. Using 14C as a reference radionuclide
the counting efficiency of PSm under different conditions was determined. Finally
a 20 mL plastic liquid scintillation vial with the lowest detection limit was selected for the activity analysis of 14C solution. [Results]: Weak interference signals from chemiluminescence and photoluminescence are observed when PSm is mixed with water. This interference can be mitigated by keeping the samples in a light-protected environment. The background count rate is positively correlated with the volume of PSm
with larger volumes resulting in higher background counts. The counting efficiency is positively correlated with the height of PSm
with higher heights corresponding to higher counting efficiency. Differences between plastic and glass liquid scintillation vials of comparable specifications are minimal. The minimum detectable activity (MDA) of the 20 mL vial is lower. Analysis of 14C solutions in the range of 574.5 to 3825.6Bq·L-1 shows good agreement between measured and expected values. [Conclusions]: The study demonstrates that the background count rate and counting efficiency of plastic scintillation microspheres are influenced by various factors such as sample volume. Under fixed conditions
PSm measurements exhibit good accuracy. PSm shows promising potential in radionuclide analysis.