戴振晖, 招什武, 杨耕, 等. 瓦里安扩充型动态楔形板的蒙特卡罗模拟与验证[J]. 核技术, 2017,40(12):120302
Zhenhui DAI, Shiwu ZHAO, Geng YANG, et al. Monte Carlo simulation and validation of the enhanced dynamic wedge of Varian accelerator[J]. Nuclear techniques, 2017, 40(12): 120302
戴振晖, 招什武, 杨耕, 等. 瓦里安扩充型动态楔形板的蒙特卡罗模拟与验证[J]. 核技术, 2017,40(12):120302 DOI: 10.11889/j.0253-3219.2017.hjs.40.120302.
Zhenhui DAI, Shiwu ZHAO, Geng YANG, et al. Monte Carlo simulation and validation of the enhanced dynamic wedge of Varian accelerator[J]. Nuclear techniques, 2017, 40(12): 120302 DOI: 10.11889/j.0253-3219.2017.hjs.40.120302.
基于蒙特卡罗方法模拟Varian直线加速器扩充型动态楔形板,建立动态楔形板的质量保证新工具。使用BEAMnrc的DYNJAWS组件模拟动态楔形板,建立加速器源模型,用DOSXYZnrc程序计算体模内的深度剂量和离轴剂量,获得不同角度楔形板的楔形因子和离轴剂量分布曲线,与相同条件的测量结果进行对比,验证动态楔形板模型的精确性。蒙特卡罗模拟结果与测量结果取得了很好的一致,偏差小于2%。表明使用蒙特卡罗程序建立扩充型动态楔形板模型是可行的,能够满足临床应用的质量保证需要。
Background
2
Enhanced dynamic wedge (EDW) is commonly used to achieve dose uniformity in the target volume in radiotherapy. However
EDW must be verified before implementation in the treatment planning system.
Purpose
2
This paper is to simulate the EDW of Varian linear accelerator by Monte Carlo method
and to construct a new tool for quality assurance of EDW.
Methods
2
EDW is simulated with the DYNJAWS module of BEAMnrc to build the source model. Depth dose and off-axis dose are calculated by use of DOSXYZnrc in a phantom to generate the EDW factor and profiles of EDW at different angles. The model is validated based on the comparison of Monte Carlo simulation and real measurement results.
Results
2
The simulation results are in good agreement with measurement results
and the deviation is less than 2%.
Conclusion
2
It demonstrates that the EDW model based on Monte Carlo simulation is effective
which would achieve the requirement of quality assurance of EDW.
P W Koken , S Heukelom , J P Cuijpers , . On the practice of the clinical implementation of enhanced dynamic wedges . Medical Dosimetry , 2003 . 28 ( 1 ): 13 - 19 . DOI: 10.1016/S0958-3947(02)00140-1 http://doi.org/10.1016/S0958-3947(02)00140-1 .
P Alaei , P D Higgins . Performance evaluation and quality assurance of Varian enhanced dynamic wedges . Journal of Applied Clinical Medical Physics , 2006 . 7 ( 1 ): 14 - 20 . DOI: 10.1120/jacmp.2027.25367 http://doi.org/10.1120/jacmp.2027.25367 .
于 浪 , 杨 波 , 李 楠 , . 应用二维电离室矩阵进行虚拟楔形板物理参数检测 . 中华放射肿瘤学杂志 , 2015 . 24 ( 3 ): 340 - 343 . DOI: 10.3760/cma.j.issn.1004-4221.2015.03.027 http://doi.org/10.3760/cma.j.issn.1004-4221.2015.03.027 .
Lang YU , Bo YANG , Nan LI , . Dosimetry verification of virtual wedge using a two-dimensional ionization chamber array . Chinese Journal of Radiation Oncology , 2015 . 24 ( 3 ): 340 - 343 . DOI: 10.3760/cma.j.issn.1004-4221.2015.03.027 http://doi.org/10.3760/cma.j.issn.1004-4221.2015.03.027 .
Varian Oncology Systems. C-series clinac enhanced dynamic wedge implementation guide[R]. Varian, 2002:1-166.
D W O Rogers , B A Faddegon , G X Ding , . BEAM:a Monte Carlo code to simulate radiotherapy treatment units . Medical Physics , 1995 . 22 ( 5 ): 503 - 524 . DOI: 10.1118/1.597552 http://doi.org/10.1118/1.597552 .
熊 绮丽 , 石 勇 , 徐 刚 , . TrueBeam加速器多叶准直器射野剂量学特性 . 核技术 , 2015 . 38 ( 10 ): 100203 DOI: 10.11889/j.0253-3219.2015.hjs.38.100203 http://doi.org/10.11889/j.0253-3219.2015.hjs.38.100203 http://www.j.sinap.ac.cn/hejishu/CN/abstract/abstract493.shtml .
Qili XIONG , Yong SHI , Gang XU , . Dosimetry characteristics of multi-leaf collimator field for TrueBeam . Nuclear Techniques , 2015 . 38 ( 10 ): 100203 DOI: 10.11889/j.0253-3219.2015.hjs.38.100203 http://doi.org/10.11889/j.0253-3219.2015.hjs.38.100203 http://www.j.sinap.ac.cn/hejishu/CN/abstract/abstract493.shtml .
M B Kakakhel , E S Baveas , A L Fielding , . Validation and automation of the DYNJAWS component module of the BEAMnrc Monte Carlo code . Australasian Physical and Engineering Sciences in Medicine , 2011 . 34 ( 1 ): 83 - 90 . DOI: 10.1007/s13246-011-0060-x http://doi.org/10.1007/s13246-011-0060-x .
F Verhaegen , H H Liu . Incorporating dynamic collimator motion in Monte Carlo simulations:an application in modelling a dynamic wedge . Physics in Medicine and Biology , 2001 . 46 ( 2 ): 287 - 296 . DOI: 10.1088/0031-9155/46/2/302 http://doi.org/10.1088/0031-9155/46/2/302 .
陈 亮 , 李 玉民 , 徐 红 . 瓦里安加速器扩充型动态楔形板楔形因子特性研究 . 中华放射肿瘤学杂志 , 2004 . 13 ( 1 ): 53 - 57 . DOI: 10.3760/j.issn:1004-4221.2004.01.013 http://doi.org/10.3760/j.issn:1004-4221.2004.01.013 .
Liang CHEN , Yumin LI , Hong XU . Characteristics of wedge factor of an enhanced dynamic wedge on Varian 600C accelerator . Chinese Journal of Radiation Oncology , 2004 . 13 ( 1 ): 53 - 57 . DOI: 10.3760/j.issn:1004-4221.2004.01.013 http://doi.org/10.3760/j.issn:1004-4221.2004.01.013 .
J P Gibbons . Calculation of enhanced dynamic wedge factors for symmetric and asymmetric photon fields . Medical Physics , 1998 . 25 ( 8 ): 1411 - 1418 . DOI: 10.1118/1.598313 http://doi.org/10.1118/1.598313 .
S A Oh , S K Kim , M K Kang , . Dosimetric verification of enhanced dynamic wedges by a 2D ion chamber array . Journal of the Korean Physical Society , 2013 . 63 ( 11 ): 2215 - 2219 . DOI: 10.1118/1.4814791 http://doi.org/10.1118/1.4814791 .
A W Beavis , S J Weston , V J Whitton . Implementation of the Varian EDW into a commercial RTP system . Physics in Medicine and Biology , 1996 . 41 ( 9 ): 1691 - 1704 . DOI: 10.1088/0031-9155/41/9/009 http://doi.org/10.1088/0031-9155/41/9/009 .
M Ahmad , J Deng , M W Lund , . Clinical implementation of enhanced dynamic wedges into the pinnacle treatment planning system:Monte Carlo validation and patient-specific QA . Physics in Medicine and Biology , 2009 . 54 ( 2 ): 447 - 465 . DOI: 10.1088/0031-9155/54/2/018 http://doi.org/10.1088/0031-9155/54/2/018 .
J H Moeller , D D Leavitt , E Klein . The quality assurance of enhanced dynamic wedges . Medical Dosimetry , 1997 . 22 ( 3 ): 241 - 246 . DOI: 10.1016/S0958-3947(97)00026-5 http://doi.org/10.1016/S0958-3947(97)00026-5 .