浏览全部资源
扫码关注微信
中国科学院上海应用物理研究所 嘉定园区 上海 201800
[ "顾颂琦, 女, 1984年出生, 2017年于上海交通大学获硕士学位, 研究领域为同步辐射实验站电子学技术", "GU Songqi, female, born in 1984, graduated from Shanghai Jiao Tong University with a master's degree in 2017, focusing on synchrotron radiation beamline technology" ]
李炯, E-mail:lijiong@sinap.ac.cn LI Jiong, E-mail:lijiong@sinap.ac.cn
收稿日期:2017-05-19,
修回日期:2017-06-13,
纸质出版日期:2017-10
移动端阅览
顾颂琦, 段佩权, 张硕, 等. 基于LabVIEW的高能量分辨谱仪数据采集系统[J]. 核技术, 2017,40(10):100102
Songqi GU, Peiquan DUAN, Shuo ZHANG, et al. A data acquisition system of high energy resolution spectrometer based on LabVIEW[J]. Nuclear techniques, 2017, 40(10): 100102
顾颂琦, 段佩权, 张硕, 等. 基于LabVIEW的高能量分辨谱仪数据采集系统[J]. 核技术, 2017,40(10):100102 DOI: 10.11889/j.0253-3219.2017.hjs.40.100102.
Songqi GU, Peiquan DUAN, Shuo ZHANG, et al. A data acquisition system of high energy resolution spectrometer based on LabVIEW[J]. Nuclear techniques, 2017, 40(10): 100102 DOI: 10.11889/j.0253-3219.2017.hjs.40.100102.
在上海光源BL14W1线站,成功搭建一套基于LabVIEW的高能量分辨谱仪运动控制和数据采集系统,该光谱仪系统可以实现高能量分辨荧光探测X射线吸收谱(High Energy Resolution Fluorescence Detected X-ray Absorption Spectroscopy,HERFD-XAS)、X射线发射谱(X-ray Emission Spectroscopy,XES)和共振X射线发射谱(Resonant X-ray Emission Spectroscopy,RXES)等主要实验功能,分辨率达到亚电子伏。硬件系统采用上海光源统一的VME运动控制硬件系统和硅漂移固体探测器;软件系统采用LabVIEW编写用户操作界面,利用DSC(Data logging and Supervisory Control Module)模块实现与运动控制使用的实验与物理工业控制系统(Experimental and Physics Industrial Control System,EPICS)系统的数据交换。利用基于该运动控制和数据采集系统的光谱仪,开展了Mn化合物和ThO
2
锕系化合物的实验。结果表明,该系统可以满足高能量分辨率实验的需求。
Background
2
High energy resolution spectrometer can perform high energy resolution fluorescence detected X-ray absorption spectroscopy (HERFD-XAS)
X-ray emission spectroscopy (XES)
resonant X-ray emission spectroscopy (RXES)
and other experiments. The energy resolution could be sub-eV level and the system can get more fine electronic structure information compared with the conventional spectral experimental methods.
Purpose
2
This study aims to develop a data acquisition system of high energy resolution spectrometer based on LabVIEW at BL14W1 X-ray absorption fine structure (XAFS) station in Shanghai synchrotron radiation facility (SSRF).
Methods
2
The hardware system includes two parts:VME motion control system for multi-dimension adjusting and silicon drift detector (SDD) fluorescence detector. The software architecture of the spectrometer consists of two parts
the experimental and physics industrial control system (EPICS) for motor control system and LabVIEW for data acquisition system (DAQ). LabVIEW's data logging and supervisory control (DSC) module is used for data exchange between the motion control system and the DAQ.
Results
2
The energy resolution of this spectrometer is 0.5 eV
which can distinguish different compounds in the mixture.
Conclusion
2
It is proved by experiment that the energy resolution and counting rate of this high energy resolution spectrometer has reached the level of the same kind of devices in the world.
A Boubnov , H W P Carvalho , D E Doronkin . Selective catalytic reduction of no over Fe-ZSM-5:mechanistic insights by operando HERFD-XANES and valence-to-core X-ray emission spectroscopy . Journal of the American Chemical Society , 2014 . 136 ( 37 ): 13006 - 13015 . DOI: 10.1021/ja5062505 http://doi.org/10.1021/ja5062505 .
R Nomura , H Ozawa , S Tateno . Spin crossover and iron-rich silicate melt in the earth's deep mantle . Nature , 2011 . 473 ( 7346 ): 199 - 202 . DOI: 10.1038/nature09940 http://doi.org/10.1038/nature09940 .
K M Lancaster , M Roemelt , P Ettenhuber . X-ray emission spectroscopy evidences a central carbon in the nitrogenase iron-molybdenum cofactor . Science , 2011 . 334 ( 6058 ): 974 - 977 . DOI: 10.1126/science.1206445 http://doi.org/10.1126/science.1206445 .
S M Butorin , K O Kvashnina , J R Vegelius . High-resolution X-ray absorption spectroscopy as a probe of crystal-field and covalency effects in actinide compounds . Proceedings of the National Academy of Sciences , 2016 . 113 ( 29 ): 8093 - 8097 . DOI: 10.1073/pnas.1601741113 http://doi.org/10.1073/pnas.1601741113 .
M H Jiang , X Yang , H J Xu , . Shanghai synchrotron radiation facility . Chinese Science Bulletin , 2009 . 54 ( 22 ): 4171 - 4181 . DOI: 10.1007/s11434-009-0689-y http://doi.org/10.1007/s11434-009-0689-y .
H S Yu , X J Wei , J Li , . The XAFS beamline of SSRF . Nuclear Science and Techniques , 2015 . 26 ( 5 ): 050102 DOI: 10.13538/j.1001-8042/nst.26.050102 http://doi.org/10.13538/j.1001-8042/nst.26.050102 .
X Gao , S Gu , Q Gao , . A high-resolution X-ray fluorescence spectrometer and its application at SSRF . X-ray Spectrometry , 2013 . 42 ( 6 ): 502 - 507 . DOI: 10.1002/xrs.2719 http://doi.org/10.1002/xrs.2719 .
P Q Duan , S Q Gu , H J Cao , . A three-crystal spectrometer for high-energy resolution fluorescence-detected X-ray absorption spectroscopy and X-ray emission spectroscopy at SSRF . X-ray Spectrometry , 2017 . 46 ( 1 ): 12 - 18 . DOI: 10.1002/xrs.2719 http://doi.org/10.1002/xrs.2719 .
W H Jia , S W Ma , L F Zheng . The control system for water-cooled DCMS in SSRF . Nuclear Science and Techniques , 2015 . 26 ( 2 ): 22 - 26 . DOI: 10.13538/j.1001-8042/nst.26.020103 http://doi.org/10.13538/j.1001-8042/nst.26.020103 .
顾 俊俊 , 邹 杨 , 魏 向军 . BLl4W1光束线XAFS实验站极低浓度数据采集系统 . 核技术 , 2011 . 34 ( 5 ): 341 - 344 .
Junjun GU , Yang ZOU , Xiangjun WEI . The low concentration data acquisition system for BLl4W1 beamline XAFS experimental station . Nuclear Techniques , 2011 . 34 ( 5 ): 341 - 344 .
邹 杨 , 顾 颂琦 , 姜 政 , . 基于LabVIEW的XAFS光束线实验站数据采集系统 . 核技术 , 2009 . 32 ( 4 ): 246 - 250 . http://www.cnki.com.cn/Article/CJFDTOTAL-HJSU200904004.htm .
Yang ZOU , Songqi GU , Zheng JIANG . Data acquisition system of XAFS beamline experimental station based on LabVIEW . Nuclear Techniques , 2009 . 32 ( 4 ): 246 - 250 . http://www.cnki.com.cn/Article/CJFDTOTAL-HJSU200904004.htm .
0
浏览量
26
下载量
3
CSCD
关联资源
相关文章
相关作者
相关机构
京公网安备11010802024621