韩俊杰, 朱金汉, 张白霖, 等. 医用直线加速器主要部件对X射线能谱的影响[J]. 核技术, 2015,38(10):100202
HAN Junjie, ZHU Jinhan, ZHANG Bailin, et al. Effects of main components of clinical linac on X-ray energy spectra[J]. Nuclear techniques, 2015, 38(10): 100202
韩俊杰, 朱金汉, 张白霖, 等. 医用直线加速器主要部件对X射线能谱的影响[J]. 核技术, 2015,38(10):100202 DOI: 10.11889/j.0253-3219.2015.hjs.38.100202.
HAN Junjie, ZHU Jinhan, ZHANG Bailin, et al. Effects of main components of clinical linac on X-ray energy spectra[J]. Nuclear techniques, 2015, 38(10): 100202 DOI: 10.11889/j.0253-3219.2015.hjs.38.100202.
研究了医用直线加速器的主要部件对X射线能谱的影响。利用蒙特卡罗软件包EGSnrc/BEAMnrc针对Varian600C医用直线加速器,模拟完整及分别去除初级准直器、均整器和次级准直器条件下的相空间文件,通过相空间文件分析程序Beamdp对相空间文件进行分析,分别得到相应条件下的X射线能谱。均整器对能谱的影响非常大,对比于无均整器的情况,有均整器时,平均能量上升了48.9%;但初级准直器、次级准直器对能谱的影响小,平均能量的变化小于1%。在射野内,对比于无均整器的情况,有均整器时,能量注量降低了52.4%,但其空间分布变均匀,非平坦度从15.09%变为1.82%;对比于无初级准直器的情况,有初级准直器时,能量注量上升了1.58%;而对比于无次级准直器的情况,有次级准直器时,能量注量下降了2.00%。均整器对X射线的能谱的影响起主要作用,它能够使X射线的平均能量上升和射野内能量注量的分布变均匀。初级准直器对X射线能谱影响小但对X射线的散射不可忽略,而次级准直器对X射线能谱的影响小而且对X射线的散射作用可以忽略,这些结果对加速器X射线束流模型的建立有一定的参考意义。
Background: It is well known that the main components of clinical linac such as the primary collimator
flattening filter and secondary collimator affect the X-ray energy spectra. How does it appear is the foundation of dose calculation
and it is helpful for building up a beam model. Purpose: This paper aims to make a detailed study on the effects of these main components on the X-ray energy spectra of a clinical linac beam. Methods: Monte Carlo program EGSnrc/BEAMnrc is employed to build up the beam model of Varian 600C linac
as well as the other three different beam models: without primary collimator
without flattening filter
without secondary collimator at the same time. After respective simulations
we use Beamdp program to analyze the phase space files
so as to obtain energy spectra for different models. Results: Flattening filter has heavy influence on X-ray beam’s energy spectra
the mean energy increases by 48.9% compared with its absence. But the primary collimator and secondary collimator have little influence on energy spectra
the change of mean energy is within 1% compared their absence. In the center field size of 10 cm×10 cm
flattening filter could let the energy fluence decrease by 52.4%
but the distribution becomes more flat and the unflatness changes from 15.09% to 1.82%; primary collimator could increase energy fluence by 1.58%; but secondary collimator could decrease the energy fluence by 2.00%. Conclusion: Flattening filter has mainly influence on the energy spectra of X-ray beam
it could let the mean energy increase and the energy fluence distribution become more flat; the scatter of primary collimator for X-ray beam must be taken into account whilst the scatter of secondary collimator for X-ray beam can be neglected. The results are meaningful to build a X-ray beam model for clinical linac. Abstract