LI Weinan, YANG Chaowen, ZHOU Rong. Baseline estimation method of digital multi-channel pulse height analyzer based on FPGA[J]. Nuclear techniques, 2015, 38(6): 060403
LI Weinan, YANG Chaowen, ZHOU Rong. Baseline estimation method of digital multi-channel pulse height analyzer based on FPGA[J]. Nuclear techniques, 2015, 38(6): 060403 DOI: 10.11889/j.0253-3219.2015.hjs.38.060403.
Background: Histogram statistics method and average value method are widely used to obtain the baseline of nuclear signal processing. However
there still exist several shortcomings
such as low accuracy
weak real-time character and high hardware resource occupancy. Purpose: This study aims to improve the accuracy and the real-time performance of the baseline extraction for the digital nuclear instruments. Methods: A novel method of obtaining baseline is proposed
it is a real-time dynamic average method that can be easily implemented on a Field-Programmable Gate Array (FPGA) based multi-channel pulse height analyzer. By setting appropiate parameters
the accuracy and real-time character could be obviously improved. Results: This novel method not only occupies less hardware resource
but also achieves higher accuracy. The energy resolution of 137Cs spectrum is got improved to 7.44%. Conclusion: This method can be applied to extract baseline of digital multi-channel pulse height analyzer with high accuracy and real-time response. Abstract