近年来,随着核辐射成像技术的发展及应用的需求,核辐射成像设备在高灵敏度、高分辨率、高信噪比以及携带方便等方面受到越来越多的关注,编码孔径准直器代替传统的平行孔和针孔准直器顺应了这种发展趋势,成为核辐射成像领域的研究热点。本文主要介绍一种新型便携式辐射成像探测系统中修正均匀冗余阵列(Modified Uniformly Redundant Arrays
MURA)编码孔径准直器的成像方法,研究其解码过程、校正方法和解码矩阵的算法:δ解码算法和精细采样平衡解码算法。在MATLAB环境下对两种解码算法进行了对比仿真,并对单个放射源与多个放射源情况分别进行了图像重建实验。实验结果表明,精细采样算法要比δ解码算法更能使成像系统对单源或多源进行清晰成像。
Background: The coded aperture collimator becomes a hot research topic in the field of nuclear radiation imaging. Purpose: This paper mainly introduces a Modified Uniformly Redundant Arrays (MURA) based on new portable radiation imaging system and focuses on the decoding process of MURA coded aperture collimator imaging system
near field correction
etc. Methods: Two kinds of decoding matrix algorithms named: δ decoding algorithm and finely sampling balance decoding (FSBD) algorithm are investigated in details. Comparison between the two decoding algorithms was conducted by MATLAB simulation
and imaging experiments using both a single and multiple point sources. Results: The experimental results show that the FSBD algorithm outperforms the δ decoding algorithm
and the imaging system represents clear reconstructed images of multiple point sources with correct position of radiation sources. Conclusion: FSBD algorithm is a better decoding approach for radiation imaging system using MURA coded aperture collimator.Cited