SHEN Wanzeng, YAN Hongbin, GAO Daqing, et al. Simulation and implementation of scanning magnet power supply[J]. Nuclear techniques, 2014, 37(2): 020402
SHEN Wanzeng, YAN Hongbin, GAO Daqing, et al. Simulation and implementation of scanning magnet power supply[J]. Nuclear techniques, 2014, 37(2): 020402 DOI: 10.11889/j.0253-3219.2014.hjs.37.020402.
Background: The scanning power for osteoporosis therapy device is required to be with high voltage output of great precision. Purpose: The design of a high voltage scanning magnet power supply is presented. Methods: Two H bridge series made the output voltage of the power supply as high as 700 V. The closed loop control strategy was adopted to produce continuous trigonometric wave or sine wave current waveform of good quality. The design idea of power supply was introduced
and then the simulation of main circuit and Pulse Width Modulation (PWM) control circuit was provided using Simplorer software. Results: Power supply design and test results were reported. The tracking error was about ±0.2%
and all the goals of this power design are realized. Conclusion: The actual tuning results of the power supply show that the design of the power supply is accomplished.Cited