CHEN Lihong, LI Yongping, ZHAO Cuilan, et al. Design of nuclear imaging system based on MURA coded aperture collimator[J]. Nuclear techniques, 2013, 36(8): 080402
CHEN Lihong, LI Yongping, ZHAO Cuilan, et al. Design of nuclear imaging system based on MURA coded aperture collimator[J]. Nuclear techniques, 2013, 36(8): 080402 DOI: 10.11889/j.0253-3219.2013.hjs.36.080402.
Background: Location nuclear imaging systems have been used in many fields
such as astronomical observation
radioactive source monitoring
nuclear development
biomedicine
etc.Purpose: It puts forward much higher requirements of monitoring and control of safety of nuclear facilities and radioactive sources.In this paper
a new portable nuclear imaging system is designed
which can be used in far field imaging.Methods: It uses high resolution detector which consists of LaBr3 crystal and H8500 position sensitive photomultiplier to meet the design requirements of high resolution.In order to acquire the higher radiation collection efficiency
we choose modified uniformly redundant arrays(MURA) as system collimator.It uses the maximum likelihood-expectation maximization(MLEM) algorithms in reconstruction procedure.Results: The test results show that the imaging system can image the far field radioactive source clearly.The field-of-view of our system can achieve the design goal: ±10° ±20°