[1]关祖杰,余君岳,杨健明.一个单采样头的支气管剂量计[J].核技术,1998(12):725-730.
GUAN Zujie. A bronchial dosimeter with a single sampler[J]. Nuclear techniques, 1998, (12): 725-730.
提出了一个只用一个采样头的便携式支气管剂量计。采样头分别由一层400目金属筛网和滤膜组成。采样气流面速为12cm/s,采样后分别对筛网和滤膜同时测量其总α计数,可得氡子体照射气管支气管(T-B区)的年有效剂量(mSv/a)。如将采样头与主动式活性炭盒串接采样测定氧气浓度,则还可以计算出以μSv/(aBqm-3)为单位的剂量转换因子。给出了这种支气管剂量计的设计思想和试验性测量结果。
A new bronchial dosimeter has been proposed which consists of only a single sampler and employs only one 400 mesh wire screen and one filter. During sampling air is drawn through the screen and filter at a face velocity of 12.0cm/s. After sampling the measurements of the screen and filter are made
respectively
by gross alpha counting. The annual effective dose (AED
in mSv/a) is then calculated in the tracheobronchial (T-B) region exposed by radon progenies. The dosimeter can be used in conjunction with an activated charcoal canister for measurements of the radon gas concentration to yield the effective dose conversion factor (DCF
in mSv/a per Bq/m3). The validity of the dosimeter has been tested through measurements of the AED and DCF under different environments. A design of the portable bronchial dosimeter is given.