A 99mTc linear source was made with a plastic cover
and SPECT/CT imaging was performed with a GE Hawkeye hybrid system and patient’s beds under load or nothing.The two kinds of data acquired were reconstructed and fused
and the deviation was quantitatively analyzed by the profiling method.Without load on the patient’s bed
the SPECT and CT images matched perfectly
with the SPECT image center being deviated from that of the CT image at only(1.37±0.96) mm.Under 25-kg load
the deviation of SPECT and CT image centers was(2.21±1.32) and(3.26±1.52) mm for new and old beds
respectively
with the statistical difference of the deviations being P<0.05 between the new and old beds.For the new and old beds under the load or nothing
the deviation of SPECT and CT image centers was significant
P<0.05
with the largest deviation being 6.0 mm.These indicate that the position change of the patient’s bed under load could result in deviation of the SPECT and CT image match