LIU Feng WANG Yugang DU Guanghua XUE Jianming WANG Sixue YAN Sha ZHAO Weijiang. STM-HOPG counter for low-energy heavy particles: principle and application[J]. Nuclear techniques, 2002, (7): 487-492.
LIU Feng WANG Yugang DU Guanghua XUE Jianming WANG Sixue YAN Sha ZHAO Weijiang. STM-HOPG counter for low-energy heavy particles: principle and application[J]. Nuclear techniques, 2002, (7): 487-492.DOI:
Based on the scanning tunneling microscope (STM) observation on energetic-particle induced damage on the surface of highly oriented pyrolytic graphite (HOPG)
a novel energetic-particle counter has been developed. The detection efficiency of the detector for heavy ions with keV energies is near 1. The count range is between 0.1×10 9-1×10 9ions·cm -2 and 0.1×10 14 -1×10 14 ions·cm -2 . With the atomic-resolution of STM
the counter can study the distribution of energetic particles in nano-scale. It can be used to detect the penetration behavior of low-energy ions through thin solid target.