WEI Yingguang1 LIU Senlin2 SUN Lifeng3 WANG Yanfeng1 1. Calculation and experimental validation of neutron response for a high sensitivity environmental neutron dose equivalent meter[J]. Nuclear techniques, 2007, (4): 376-380.
WEI Yingguang1 LIU Senlin2 SUN Lifeng3 WANG Yanfeng1 1. Calculation and experimental validation of neutron response for a high sensitivity environmental neutron dose equivalent meter[J]. Nuclear techniques, 2007, (4): 376-380.DOI:
The energy response of a high sensitivity environmental neutron dose equivalent meter is calculated by MCNP4c code. The calculated response curve agrees with H *(10) curve in ICRP 74 report in interesting neutron energy regions. By comparison between the experimental results and the calculated values
the relative deviation is -21.6%. Experiments show that MCNP4C code can be applied to calculate neutron response for the high sensitivity environmental neutron dose equivalent meter
and it can also raise the work efficiency and help to analyze the experimental results.