ZHANG Jiangyun1 HUANG Ning1 TANG Lili 2 LIU Yanfang2 ZHANG Guanghua3 1. Application of Monte Carlo method in determination of air absorbed dose rate from a full γ-ray spectrum[J]. Nuclear techniques, 2011, 34(8): 585-588.
ZHANG Jiangyun1 HUANG Ning1 TANG Lili 2 LIU Yanfang2 ZHANG Guanghua3 1. Application of Monte Carlo method in determination of air absorbed dose rate from a full γ-ray spectrum[J]. Nuclear techniques, 2011, 34(8): 585-588.DOI:
The dosimetric properties of a gamma-radiation field can be measured by a dose-rate meter or by a gamma-spectrometer with spectrum-dose conversion.One of the spectrum dose conversion methods is to evaluate the radiation dosedirectly from integrating the observed pulse height spectrum weighted by a G(E) function
instead of making spectrum analysis.In this paper
energy spectra of 11 single energy point sources in 0.1–2.5 MeV are simulated by using Monte Carlo software MCNP5 for NaI(Tl) detectors of Φ 75 mm×75 mm
Φ 50 mm×50 mm and Φ 25 mm×25 mm.The energy spectra are used as standard spectrum to calculate the G(E) value
so as to calculate corresponding doses.Comparing the results with theoretical value
the error is less than 1.5%.Finally
the G(E)-based γ dose rate of radiation field agreed well(within 5%) with the dose rate measured by a dosimeter.