徐海珊, 向正瑜, 牟运山, 等. 一台测量快中子辐射俘获截面的大液体闪烁探测器[J]. 核技术, 1986,09(9):5-8,56.
Haishan Xu, Zhengyu Xiang, Yunshan Mu, et al. A large liquid scintillation counter for fast neutron capture cross section measurements[J]. Nuclear techniques, 1986, 09(9): 5-8,56.
为了测量快中子辐射俘获截面,我们研制了一台测量γ射线的球形液体闪烁探测器,其直径为1m,容积680 l。该探测器属4π几何类型,探测器效率达80%以上。为了降低本底,除了用10cm厚的铅和40cm厚的石蜡建成屏蔽室外,还采用符合方法,使本底计数小于100cps。探测器已经在2.5 МeV脉冲质子静电加速器上测量了金、钽和铥等元素的辐射俘获截面。
For determining fast neutron radiative capture cross sections a new capture gammaray counter of approximately 4π geometry has been developed. The counter is in a sphere shape with a diameter of 1 meter filled with 680 liter liquid scintillator. The tank in shielded by 10 cm thickness of lead and 40 cm thickness of Li-paraffin. A coincidence method is used for reducing background due to neutron capture by the hydrogen in the liquid scintillator. The counter has been successfully used for measuring the neutron capture cross sections of Au-197
Ta-181 and Tm-169 on a 2.5 MeV pulsed Van de Graaff accelerator.