A new precise instrument for determination of the absolute disintegration of β
α emitter solutions by the method of liquid scintillation counting is described. The instrument includes a triple phototube liquid scintillation detector and three output counting channels which are used for the number of triple coincidences Nc
the number of sum of double coincidences N
D
and the number of sum of single phototube counting N
E
seperately. This instrument is based on the principle of extrapolating the triple to double coincidence ratio (TDCR) to unity.
Examples of application of the instrument for
3
H and
14
C are given.
The instrument has two main advantages: 1. Higher accuracy. The overall measurement uncertainty(3σ)
for
3
H
63
Ni
35
S and
14
C etc. is in 0.5—1.5%. 2. Easy to operate. The measurement time is much less than usual metrology instrument.