本文叙述了用自己研制的组织等效电离室和组装的G-М计数管γ剂量计测定快中子比释动能的方法和测量结果,测量结果总的估计不确定度为±5.2%。并与用美国FWT IC-17A组织等效电离室和FWT GM-2计数管所测的结果进行了比对,两者符合得很好。
The two detector method was applied for kerma measurements in 15 MeV D-T mixed neutron-photon field. The hydrogenous detector employed is the thimble- type IRM 81-TE1 tissue equivalent ion chamber developed by us and the non-hydrogenous detector is the neutron-insensitive philips 18509 Geiger-Mueller counter. This paper describes the experimental devices
methods and measurement results. The results of IRM 81T-TE1 ion chamber are found to be in good agreement with values obtained by the FWT IC-17A ion chamber.
陈迪 等 , 辐射防护 , 6 , 1 ( 1983 ).
Far West Technology, Inc . , Ionization chamber calibration (IC-17A) , ( 1981 ).
J. J. Broerse et al . , Brit. J. Radiology , 54 , 882 ( 1981 ).
ICRU Report 26 , 1977 .
L. J. Goodman and J. J. Coyne , Radiat. Res. , 82 , 13 ( 1980 ).
APPM Report No. 7, Protocol for neutron beam dosimetry , 1980 .
ICRU Report 31 , 1979 .
R. S. Caswell et al . , Radiat. Res. , 83 , 217 ( 1980 )
郭勇 等 , 计量学报 , 5 , 2 , 107 ( 1984 ).
J. C. McDONALD et al . Radiation protection, Fourth symposium on neutron dosimetry , 1981 , p, 515 .
J. J. Broerse ed. , Ion chambers for neutron dosimetry , Harwood Academic Publishers , 1980 , p. 61 .
ibid , p. 185 .
ibid , p. 73 .
ibid , p 31 .
ibid , p. 35 .
ibid , p. 131 .
D. Greene and R. L. Thomas , Phys. Med. Biol. , 14 , 45 ( 1969 ).
G. Burger and H. G. Ebert eds. , Proceeding third symp. neptron dosimetry in biology and medicine , Commission of the European Communities , Luxembourg , 1978 , p. 165 .
0
浏览量
26
下载量
0
CSCD
关联资源
相关文章
相关作者
相关机构
京公网安备11010802024621