Yuexing Wang, Yongjie Lu. The measuring of thermal neutron dose with a thermoluminescent dosimeter wrapped in a cadmium enclosure[J]. Nuclear techniques, 1985, 08(5): 36-39,59.
Yuexing Wang, Yongjie Lu. The measuring of thermal neutron dose with a thermoluminescent dosimeter wrapped in a cadmium enclosure[J]. Nuclear techniques, 1985, 08(5): 36-39,59.DOI:
by which the TLD (thermoluminescent detector) wrapped in a cadmium enclosure can be used to measure thermal neutron dose equivalent in environmental monitoring. To calibrate the TLD wrapped in a cadium enclosure it is not necessary to use the reactor thermal neutron source
it can be simply accomplished with a
60
Co-
γ
source.
The formula used to calculate the equivalent thermal neutron dose is given.