Changai Wang, Shulin Liu, Pinhua Jiang, et al. Experimental study on the reference filtered X radiations used for calibrating radiation protection instruments[J]. Nuclear techniques, 1990, 13(3): 153-157.
Changai Wang, Shulin Liu, Pinhua Jiang, et al. Experimental study on the reference filtered X radiations used for calibrating radiation protection instruments[J]. Nuclear techniques, 1990, 13(3): 153-157.DOI:
Study has been made on the characteristics of the reference filtered X radiations with photon energies between 30 and 250 keV and the method of producing these radiations in a secondary standard dosimetry laboratory. The radiation parameters and some examples of calibrating portable X and Gamma radiation dose ratemeters were given.
关键词
Keywords
references
International organization for standardization , ISO4037 , 1979 , p. 4 .
顾继青 等 , 计量测试技术 ,( 1981 ), 2 : 10 。
E. Storm and H. I. Israel , Nuclear Data Tables , A7 ( 1970 ), 565 .