Zhutang Wen, Riyou Luo, Lin Gao, et al. Studies on dichromate solution/Ag+ dosimeters as 60Co absorption dose transfer standard in the 10—40 kGy range[J]. Nuclear techniques, 1990, 13(1): 29-33.
Zhutang Wen, Riyou Luo, Lin Gao, et al. Studies on dichromate solution/Ag+ dosimeters as 60Co absorption dose transfer standard in the 10—40 kGy range[J]. Nuclear techniques, 1990, 13(1): 29-33.DOI:
dosimeters in the 10—40 kGy range the response to dose is in good linearity
and the effects of dose rate and temperature during irradiation on absorption dose response could be negligible. Furthermore
the dosimeter solutions have good storage stability before and after irradiation. So long as the preparation of dosimeters and the measuring procedure are well controlled
the precision of the system can reach to about 1%
and total uncertainty to ±3.5%. The system can be used as transfer standard dosimeters for high-dose measurement.