浏览全部资源
扫码关注微信
1.中国计量科学研究院
2.军事医学科学院放射医学研究所
收稿日期:1987-08-25,
纸质出版日期:1989-10
移动端阅览
张之彬, 陈迪, 郭勇. 组织等效电离室光子定度的研究及其验证[J]. 核技术, 1989,12(10):598-603.
Zhibin Zhang, Di Chen, Yong Guo. Calibration of a tissue-equivalent ionization chamber and the experimental evidence[J]. Nuclear techniques, 1989, 12(10): 598-603.
本文详细地研究了γ定度的方法,从而求得了空腔中气体的质量,以计算吸收剂量。还直接测量了空腔的体积,由气体的密度推出了空腔气体的质量。这两种方法得到的结果在0.5%内符合。这表明:本文采用的定度程序、引用的物理常数是正当的。本文介绍的测定气体质量的方法亦可作为测定(
Wc/e
)•
S
m
g
的一个实验方法。
A calibration procedure in a photon beam for a tissue-equivalent ionization chamber is described. To verify the calibration we also measure the volume of cavity and obtained the mass of gas in the cavity directiy using gas density. The agreement between the two methods is better than 0.5%
showing that the procedure and the physical parameters used in the calibration are proper. The method used for measuring the mass of gas in the cavity is also valid for the experimental determination of (
W
C
/e
) • S
m
g
.
J. R. Williams and J. R. Greening , Phys. Med. Bial. , 25 ( 1980 ), 2 : 215 .
CEC ., Ion chambers for neutron dosimetry , Harwood Acaeemic Publishers , 1980 .
J. J. Broerse et al ., Brit. J. Radiology , 54 ( 1981 ), 882 .