葛良全, 章晔. 同位素源X射线荧光仪在某铅锌矿巷壁X取样中的应用[J]. 核技术, 1989,12(1):24-31.
Liangquan Ge, Ye Zhang. Applications of the radioisotope source X-ray fluorescence analyser on outcrop of a Pb-Zn deposit[J]. Nuclear techniques, 1989, 12(1): 24-31.
本文对某铅锌矿系统地开展了同位素源X射线荧光测量,运用了校正基体效应的方法对2号矿体群的粉末样和坑道巷壁进行了研究,建立了数学模型。经实践验证,X取样结果与刻槽取样结果相比较,在规定的允许误差限内,其合格率铅样达77%,锌样达85%,且优于重复刻槽取样的精度。重复X取样的相对误差小于10%。
Radioisotope source X-ray fluorescence analysis technique has been applied to a Pb-Zn deposit. The method of correcting matrix effect has been proposed. Having studied powder samples and outcrop in tunnel for a Pb-Zn orebody group
a mathematical model has been found. It is proved that 77% of Pb samples and 85% of Zn samples for X-ray measurement sampling on outcrop are consistent with the standards from channel sampling. The pricision of X-ray measurement sampling on outcrop is less than 10%.
16mm source-sample distance is proved to be optimum
according to the experiments on an artificial "Pb-Zn ore" model in which surface is especially prepared by grinding.
A
238
Pu-
241
Am double beam source has solved the problem of correcting scattered radiation for the portable X-ray fluorescence analyser working in tunnel.
章晔 , X射线荧光探矿技术 , 北京 , 地质出版社 , 1984 年, 第155页 。
中国地球物理学会 编 , 地球物理文集 , 中国地球物理学会出版 , 1983 年, 第156页 。
IAEA , Nuclear techniques in geochemitry and geophysics , Division of Publication , Vienna , 1976 , p. 21 .
M. J. Gallagher , Mining and petroleum geology , Applied Science Published , London , 1970 , p. 691 .
E. П. Леман , Ренгынорадатричиткий метод опробования местораждений цветных и рецких метакков, издательство , Недра , Ленинград , 1973 , ст. 143 .
章晔 等 , 成都地质学院院报 ,( 1983 ), 2 : 72 。
葛良全 , 同位素源X射线荧光法的金矿勘查及在铅锌矿上X取样的研讨 , 成都地质学院 研究生论文 , 1987 年, 第28页 。