Yan Wang, Zhenqiu Hu, Jinshui Zhou, et al. A study on small volume liquid scintillation system of radioimmunoassay[J]. Nuclear techniques, 1988, 11(2): 59-59,6,56.
Yan Wang, Zhenqiu Hu, Jinshui Zhou, et al. A study on small volume liquid scintillation system of radioimmunoassay[J]. Nuclear techniques, 1988, 11(2): 59-59,6,56.DOI:
Our experiments show that the H-number quench correction method may be used not only in standard liquid scintillation counting but also in small volume liquid scintillation counting. Considering this characteristic and the sample holding capacity
the RIA' sample volume is reduced and poured into a disposable polyethylene tube with cap (12×50mm) containing one milliliter scintillation cocktail for counting. It is more convenient for operating and more than 80% of scintillation cocktail can be saved. The system has been used in RIA of tetosterone
estradiol
estriol
chloropromazine and so on. The results are reliable.
关键词
Keywords
references
D. L. Horrocks , Counting Exchange , 1 ( 1985 ), 1 : 1 .