The fabrication of a single-ended coaxial Ge(Li) detector
工作简报|更新时间:2022-08-16
|
The fabrication of a single-ended coaxial Ge(Li) detector
同轴单开端型Ge (Li) 探测器的研制
Nuclear Techniques1981年04卷第3期
作者机构:
作者简介:
基金信息:
DOI:
中图分类号:
收稿日期:1980-05-23,
纸质出版日期:1981-06
稿件说明:
移动端阅览
Group of Ge (Li) Detector, Institute of Nuclear Energy Technology, Qinghua University. The fabrication of a single-ended coaxial Ge(Li) detector[J]. Nuclear Techniques, 1981,04(3):46-48.
Group of Ge (Li) Detector, Institute of Nuclear Energy Technology, Qinghua University. The fabrication of a single-ended coaxial Ge(Li) detector[J]. Nuclear Techniques, 1981,04(3):46-48.DOI:
In this paper the fabrication technique of a single-ended coaxial Ge (Li) detector is described. The active volume of the detector is about 35cm3. The leakage current at 2500V bias is 4×10
-11
A. An energy resolution of 2.47 keV (FWHM) for
60
Co (1.332 MeV) has been obtained.
The technique of etch of surface without protection and the method of clean-up drift in the cryostat are discussed in detail.