The measurement of absolute intensity of medical isotopes 131I and 198Au by efficiency extrapolation method
|更新时间:2022-08-16
|
The measurement of absolute intensity of medical isotopes 131I and 198Au by efficiency extrapolation method
用效率外推法测131I和198Au绝对强度
Nuclear Techniques1981年04卷第1期
作者机构:
Shanghai Institute of Nuclear Research, Academia Sinica
作者简介:
基金信息:
DOI:
中图分类号:
收稿日期:1979-12-26,
纸质出版日期:1981-02
稿件说明:
移动端阅览
Zhu Jia-bi, Liu Jing-yi, Lü Shang-nü, 等. The measurement of absolute intensity of medical isotopes 131I and 198Au by efficiency extrapolation method[J]. Nuclear Techniques, 1981,04(1):16-20,30,30.
Zhu Jia-bi, Liu Jing-yi, Lü Shang-nü, 等. The measurement of absolute intensity of medical isotopes 131I and 198Au by efficiency extrapolation method[J]. Nuclear Techniques, 1981,04(1):16-20,30,30.DOI:
Au with a short half-life has been measured by the efficiency extrapolation method of 4
πβ-γ
coincidence absorption curve. The result of the measurement has been corrected for background
dead time
resolving time respectively. The correction for the short half-life has also been considered. The maximum error of the measurement are 0.57% and 1.7% for
131
I and
198
Au respectively. After a little change of the apparatus
we have measured the nuclide with electron capture decay.