Applying the statistical theory of radioactive disintegration and the formula of error transmission. We have derived the expressions for the total standard error of dpm values obtained in liquid scintillation counting after quench correction. By the aid of an electronic computer
a series of tables for the control of errors in measurement have been obtained. Using these tables
one may conveniently control errors of his samples in measurement.