A large area gas-flow counter——design, construction and operational method
|更新时间:2022-08-16
|
A large area gas-flow counter——design, construction and operational method
大面积流气式计数管的设计、制造和使用方法
Nuclear Techniques1980年03卷第4期
作者机构:
作者简介:
基金信息:
DOI:
中图分类号:
收稿日期:1979-10-05,
纸质出版日期:1980-08
稿件说明:
移动端阅览
Radiation Counter Group, Institute of Biophysics, Academia} Sinica. A large area gas-flow counter——design, construction and operational method[J]. Nuclear Techniques, 1980,03(4):15-18.
Radiation Counter Group, Institute of Biophysics, Academia} Sinica. A large area gas-flow counter——design, construction and operational method[J]. Nuclear Techniques, 1980,03(4):15-18.DOI:
A flat type large area gas-flow counter with multi-wire anode was constructed in 1971 to monitor beta active waste with low specific activity directly. The design consideration
material
process of manufacture
operational method and technical performance of this counter are described.
The sensitive area of main counter is up to 1000cm
2
and that of anticoincidence counter is 2500cm
2
. With a 10cm thick lead shield the background count rate is less than 25cpm (specific background count rate is less than 0.025cpm/cm