The study on the gain of microchannel plate
NUCLEAR TECHNIQUESVol. 42, Issue 10, (2019)
作者机构:
1.西北核技术研究院 西安 710024
2.强脉冲辐射环境模拟与效应国家重点实验室 西安 710024
作者简介:
ZHANG Xiaodong, male, born in 1981, graduated from Northwest Institute of Nuclear Technology with a doctoral degree in 2013, focusing on pulsed radiation measurement
基金信息:
National Natural Science Foundation of China(11375142)
ZHANG Xiaodong,OUYANG Xiaoping,HE Junzhang,et al.The study on the gain of microchannel plate[J].NUCLEAR TECHNIQUES,2019,42(10):46-50. DOI: 10.11889/j.0253-3219.2019.hjs.42.100403.
Microchannel plate (MCP) is an electron multiplier device with an ultra-fast time response
which can be directly used for the detection of charged particles. The gain of MCP has a large influence on the properties of the neutron detector.
Purpose
2
This study aims to investigate the gain of the MCP through theoretical simulation and experimental test.
Methods
2
First of all
a sub-ns time-response neutron detector based on recoiled proton and MCP was developed for experimental test.The Monte Carlo method was used to simulate the number of secondary electrons generated by a single electron multiplying on a single microchannel
and the electron gain of a single-chip MCP at different voltages was calculated. Then
the gain characteristics of two MCPs were studied by alpha particle experiments.
Results
2
The calculation results show that the theoretical gain of a single-chip MCP is consistent with the MCP factory inspection report. The experimental results show that when the supply voltage of each MCP is 800 V
the gain of double MCP is about 9.0×10
6
.
Conclusions
2
The method clarified in the paper is reliable and feasible
and provides a new technical approach for the study of MCP gain. It also provides a theoretical basis for studying the performance of MCP neutron detectors.
PAN Jingsheng . The developments of low noise high resolution microchannel plate [D ] . Nanjing : Nanjing University of Scienee and Technology , 2008 : 5 ‒ 8 .
XU Shenyue , MA Xinwen , YAN Shuncheng , et al . Multi hit three layer delay line anode position sensitive detector based on microchannel plate [J ] . Nuclear Electronics & Detection Technology , 2009 , 29 ( 1 ): 96 ‒ 100 .
SUN Lijuan , CUI Mingqi , ZHU Jie , et al . Absolute calibration of microchannel plates in soft X-ray region [J ] . Nuclear Techniques , 2007 , 30 ( 5 ): 414 ‒ 418 .
ZHANG Xiaodong . Technology of the ultrafast pulse neutron detector based on recoil proton and MCP [R ] . Xi'an: Northwest Institute of Nuclear Technology , 2017 : 11 ‒ 30 .
PENG Lingling . Investigation of microchannel plate film dynode and secondary electron emission characteristics [D ] . Changchun: Changchun University of Scienee and Technology , 2014 .
卢其亮 . 高速离子诱发二次电子发射的模拟研究 [D ] . 上海 : 复旦大学 , 2002 .
LU Qiliang . Simulation of secondary electron emission induced by swift ions [D ] . Shanghai: Fudan University , 2002 .
XU Yanhui . Design and research of secondary electron yield test equipment with high precision [D ] . Hefei:University of Science and Technology of China , 2016 .