Study on the accuracy of detection efficiency for radioactive iodine monitor
NUCLEAR ELECTRONICS AND INSTRUMENTATION|更新时间:2024-10-25
|
Study on the accuracy of detection efficiency for radioactive iodine monitor
“In the field of nuclear environmental protection, experts use Monte Carlo software to simulate and calculate the detection efficiency of iodine monitoring instruments, providing reference for improving detection accuracy. The results indicate that for coaxial well type activated carbon sampling boxes, the detection efficiency and accuracy of uniform distribution source calibration are relatively high.”
NUCLEAR TECHNIQUESVol. 42, Issue 1, (2019)
作者机构:
1.北京市射线应用研究中心 北京 100015
2.北京市辐射中心 北京 100875
作者简介:
YIN Yin, female, born in 1986, graduated from China University of Geosciences (Beijing) with a master's degree in 2013, focused in nuclear instruments and instruments
YIN Yin,HUANG Zihan,XU Hongkun,et al.Study on the accuracy of detection efficiency for radioactive iodine monitor[J].NUCLEAR TECHNIQUES,2019,42(01):33-37.
YIN Yin,HUANG Zihan,XU Hongkun,et al.Study on the accuracy of detection efficiency for radioactive iodine monitor[J].NUCLEAR TECHNIQUES,2019,42(01):33-37. DOI: 10.11889/j.0253-3219.2019.hjs.42.010401.
Study on the accuracy of detection efficiency for radioactive iodine monitor