Wei CHEN, Jianbin ZHOU, Fang FANG, et al. Design of digital multi-channel spectrometer based on auto-baseline restoration[J]. Nuclear techniques, 2018, 41(5): 050401
Wei CHEN, Jianbin ZHOU, Fang FANG, et al. Design of digital multi-channel spectrometer based on auto-baseline restoration[J]. Nuclear techniques, 2018, 41(5): 050401 DOI: 10.11889/j.0253-3219.2018.hjs.41.050401.
不同放射性强度的测量应用中,探测器输出核脉冲信号的基线将发生漂移,进而造成多道能谱仪的谱线漂移和能量分辨率损失。尽管采用数字基线估计方法可以对核脉冲信号的基线进行正确估计,但纯数字基线恢复算法无法有效处理核脉冲信号基线漂移对前端电路的影响。采用最小平均值基线估计方法,先在现场可编程门阵列(Field-Programmable Gate Array,FPGA)中进行数字基线估计;然后通过SPI总线将基线估计值传入微控制器(Micro-Controller Unit,MCU),MCU根据基线值判断是否进行基线调节,将基线估计值传入数模转换器(Digital-to-analog Converter,DAC);最后在前端电路中实现核脉冲信号的基线恢复。基于LaBr
3
(Ce)探测器的测试结果表明:采用该基线自动恢复技术的数字化多道能谱仪能够实现核脉冲信号的基线自动恢复。在高放射性测量条件下,测量系统能够解决因基线漂移引起的谱线漂移问题,使系统能量分辨率保持稳定。
Background
2
The baseline drift of detector output could result in spectrum drift and energy resolution loss of the multi-channel spectrometer. Though baseline estimation approaches can obtain the baseline of nuclear signals
they are not able to reduce the effect of baseline drift on the front-end circuit.
Purpose
2
This study aims to design and implement a multi-channel spectrometer with auto-baseline restoration functionality.
Methods
2
First of all
the baseline of detector's output is estimated by using minimum average method and implemented in the field-programmable gate array (FPGA). Then the estimation value is sent to micro-controller unit (MCU) for regulation condition judgment
and to digital-to-analog converter (DAC) for necessary baseline adjustment. Finally
the baseline of detector output is restored in the front-end circuit. A multi-channel spectrometer is designed based on LaBr
3
(Ce) detector and this auto-baseline restoration functionality.
Results
2
The experimental results imply that the designed multi-channel spectrometer can realize baseline restoration automatically.
Conclusion
2
Under the situation of high radioactivity
the baseline drift of detector output can be solved by this multi-channel spectrometer
and the energy resolution of the system can remain stable.
J B Sim es , C Correia . Pulse processing architectures . Nuclear Instruments & Methods in Physics Research , 1999 . 422 ( 1-3 ): 405 - 410 . DOI: 10.1016/S0168-9002(98)00992-9 http://doi.org/10.1016/S0168-9002(98)00992-9 .
陈亮. 核素识别算法及数字化能谱采集系统研究[D]. 北京: 清华大学, 2009. http://cn.bing.com/academic/profile?id=8e7906b9edc0aed5784bcfe1b852dbcc&encoded=0&v=paper_preview&mkt=zh-cn
CHEN Liang. Nuclide identification algorithm and digital spectra acquisition system[D]. Beijing: Tsinghua University, 2009.
R Abbiati , E Gatti , A Geraci . A new digital estimation technique for baseline restoration . Nuclear Instruments & Methods in Physics Research A , 2005 . 548 ( 3 ): 507 - 516 . DOI: 10.1016/j.nima.2005.04.072 http://doi.org/10.1016/j.nima.2005.04.072 .
A Geraci , I Rech , E Gatti . Shared baseline restoration at minimum noise for high resolution spectroscopy . Nuclear Instruments and Methods in Physics Research A , 2002 . 482 ( 1-2 ): 441 - 448 . DOI: 10.1016/S0168-9002(01)01509-1 http://doi.org/10.1016/S0168-9002(01)01509-1 .
张 同锋 , 方 方 , 王 敏 , . 核辐射脉冲幅度分析的基线自适应Kalman滤波估计 . 核电子学与探测技术 , 2012 . 32 ( 2 ): 203 - 206 . DOI: 10.3969/j.issn.0258-0934.2012.02.020 http://doi.org/10.3969/j.issn.0258-0934.2012.02.020 .
Tongfeng ZHANG , Fang FANG , Min WANG . Baseline adaptive Kalman filter estimation method for nuclear radiation pulse height analysis . Nuclear Electronics & Detection Technology , 2012 . 32 ( 2 ): 203 - 206 . DOI: 10.3969/j.issn.0258-0934.2012.02.020 http://doi.org/10.3969/j.issn.0258-0934.2012.02.020 .
肖 无云 , 魏 义祥 , 艾 宪芸 . 多道脉冲幅度分析中的数字基线估计方法 . 核电子学与探测技术 , 2005 . 25 ( 6 ): 601 - 604 . https://www.wenkuxiazai.com/doc/5adb4f5277232f60ddcca1b3.html .
Wuyun XIAO , Yixiang WEI , Xianyun AI . Digital baseline estimation method for multi-channel pulse height analyzing . Nuclear Electronics & Detection Technology , 2005 . 25 ( 6 ): 601 - 604 . https://www.wenkuxiazai.com/doc/5adb4f5277232f60ddcca1b3.html .
A Geraci , A Pullia , G Ripamonti . Quasi-optimum gamma and X spectroscopy based on real-time digital techniques . Nuclear Instruments & Methods in Physics Research A , 2000 . 439 ( 2 ): 378 - 384 . DOI: 10.1016/S0168-9002(99)00897-9 http://doi.org/10.1016/S0168-9002(99)00897-9 .
李 彦波 , 李 朝海 , 何 子述 , . DMCA中基于补偿最小二乘法的基线估计算法 . 核电子学与探测技术 , 2014 . 34 ( 8 ): 991 - 994 . DOI: 10.3969/j.issn.0258-0934.2014.08.017 http://doi.org/10.3969/j.issn.0258-0934.2014.08.017 .
Yanbo LI , Chaohai LI , Zishu HE . A baseline estimation algorithm based on penalized least squares in DMCA . Nuclear Electronics & Detection Technology , 2014 . 34 ( 8 ): 991 - 994 . DOI: 10.3969/j.issn.0258-0934.2014.08.017 http://doi.org/10.3969/j.issn.0258-0934.2014.08.017 .
M Kafaee , A Moussavi-Zarandi . Baseline restoration and pile-up correction based on bipolar cusp-like shaping for high-resolution radiation spectroscopy . Journal of the Korean Physical Societ , 2016 . 68 ( 8 ): 960 - 964 . DOI: 10.3938/jkps.68.960 http://doi.org/10.3938/jkps.68.960 .
李 伟男 , 杨 朝文 , 周 荣 . 基于FPGA脉冲幅度分析器的数字化基线估计方法 . 核技术 , 2015 . 38 ( 6 ): 060403 DOI: 10.11889/j.0253-3219.2015.hjs.38.060403 http://doi.org/10.11889/j.0253-3219.2015.hjs.38.060403 http://www.j.sinap.ac.cn/hejishu/CN/abstract/abstract432.shtml .
Weinan LI , Chaowen YANG , Rong ZHOU . Baseline estimation method of digital multi-channel pulse height analyzer based on FPGA . Nuclear Techniques , 2015 . 38 ( 6 ): 060403 DOI: 10.11889/j.0253-3219.2015.hjs.38.060403 http://doi.org/10.11889/j.0253-3219.2015.hjs.38.060403 http://www.j.sinap.ac.cn/hejishu/CN/abstract/abstract432.shtml .
梁 卫平 , 胡 颖睿 , 肖 无云 , . 数字化多道脉冲幅度分析器调理电路设计 . 核电子学与探测技术 , 2012 . 32 ( 4 ): 462 - 465 . DOI: 10.3969/j.issn.0258-0934.2012.04.021 http://doi.org/10.3969/j.issn.0258-0934.2012.04.021 .
Weiping LIANG , Yingrui HU , Wuyun XIAO . Digital multi-channel pulse height analyze pre-filter circuit design . Nuclear Electronics & Detection Technology , 2012 . 32 ( 4 ): 462 - 465 . DOI: 10.3969/j.issn.0258-0934.2012.04.021 http://doi.org/10.3969/j.issn.0258-0934.2012.04.021 .
钱 建复 , 沈 庭云 . 核辐射剂量学 , : 北京 国防工业出版社 , 2009 . 161 - 193 .
Jianfu QIAN , Tingyun SHEN . Nuclear radiation dosimetry , : Beijing National Defense Industry Press , 2009 . 161 - 193 .
曹 忠 , 黄 正德 , 陈 群 . 高计数率下闪烁体-光电倍增管系统增益的稳定 . 核技术 , 1990 . 13 ( 2 ): 85 - 90 . http://www.cnki.com.cn/Article/CJFDTotal-HJSU199002003.htm .
Zhong CAO , Zhengde HUANG , Qun CHEN . Gain stabilization of the scintillator-photomultiplier counting rates . Nuclear Techniques , 1990 . 13 ( 2 ): 85 - 90 . http://www.cnki.com.cn/Article/CJFDTotal-HJSU199002003.htm .
张 思玲 , 董 成富 , 原 振东 , . 低功耗和快速补偿的光电倍增管偏置电路 . 核电子学与探测技术 , 2000 . 20 ( 3 ): 236 - 239 . https://www.wenkuxiazai.com/doc/27ee0b61783e0912a2162ac9-2.html .
Siling ZHANG , Chengfu DONG , Zhendong YUAN . Low consumption and fast recovery photomultiplier bases . Nuclear Electronics & Detection Technology , 2000 . 20 ( 3 ): 236 - 239 . https://www.wenkuxiazai.com/doc/27ee0b61783e0912a2162ac9-2.html .
魏 福利 , 王 培伟 , 袁 媛 , . 光电倍增管线性电流扩展研究 . 核电子学与探测技术 , 2011 . 31 ( 9 ): 986 - 990 . DOI: 10.3969/j.issn.0258-0934.2011.09.012 http://doi.org/10.3969/j.issn.0258-0934.2011.09.012 .
Fuli WEI , Peiwei WANG , Yuan YUAN . The study of linear current extention of PMT . Nuclear Electronics & Detection Technology , 2011 . 31 ( 9 ): 986 - 990 . DOI: 10.3969/j.issn.0258-0934.2011.09.012 http://doi.org/10.3969/j.issn.0258-0934.2011.09.012 .
0
Views
35
下载量
3
CSCD
Publicity Resources
Related Articles
Related Author
Related Institution
京公网安备11010802024621