Optimization design for energy response of the mobile radioactive gas monitor
移动式放射性惰性气体监测仪对不同能量的γ射线存在不同的能量响应,为了提高其测量γ射线的准确性,需要对该监测仪进行能量响应补偿。利用MCNP(Monte Carlo N Particle Transport Code)软件模拟监测仪中的NaI(Tl)晶体在不同铅孔模型下的能量响应,通过比较不同模型下的标准偏差,选出屏蔽铅孔的最佳尺寸。模拟得出补偿铅孔的半径为3.0 cm、厚度为0.5 cm时,该监测仪的能量响应得到较好补偿。
Abstract
Background
2
The mobile radioactive gas monitor responses differently to gamma rays at different energies.
Purpose
2
This study aims to accurately measure the dose of gamma radiation monitor by means of compensation for energy response.
Methods
2
Energy responses of NaI(Tl) crystals in the monitor in different lead-hole models were simulated using Monte Carlo N Particle Transport Code (MCNP) software. Various sizes of shielding lead hole were selected to compare the standard deviations of different models
and the optimum size of shielding lead hole was obtained.
Results & Conclusion
2
The simulation result shows that the energy response of monitor is best compensated when the radius of hole is 3.0 cm and thickness of shielding lead is 0.5 cm.
Bing ZENG , Liangsheng GE , Shengliang GUO , . Monte Carlo simulation for the response function of NaI (Tl) crystal to γ-ray . Nuclear Electronics & Detection Technology , 2009 . 29 ( 2 ): 394 - 397 . DOI: 10.3969/j.issn.0258-0934.2009.02.040 http://doi.org/10.3969/j.issn.0258-0934.2009.02.040 .
Renbo WANG , Kui YANG . Monte Carlo simulation and optimization design for NaI detector energy response . Nuclear Electronics & Detection Technology , 2015 . 35 ( 2 ): 188 - 192 . DOI: 10.3969/j.issn.0258-0934.2015.02.018 http://doi.org/10.3969/j.issn.0258-0934.2015.02.018 .
Hongyan WANG , Xuefu XIAO , Yongfu MA . Energy response characteristics of ionization chamber by using Monte-Carlo method . Atomic Energy Science and Technology , 2006 . 40 ( 6 ): 28 - 32 . DOI: 10.3969/j.issn.1000-6931.2006.06.019 http://doi.org/10.3969/j.issn.1000-6931.2006.06.019 .
ICRP. Conversion coefficients for use in radiological protection against external radiation[R]. ICRP Publication 74, Oxford: Pergamon Press, 1996. http://ci.nii.ac.jp/ncid/BA3759100X?l=en
WU Xiangyu. The simulation of response function and detection efficiency for γ-ray detector of different sizes by Monte Carlo method[D]. Chengdu: Chengdu University of Technology, 2009.
General Administration of Quality Supervision, Inspection and Quarantine of the People's Republic of China. JJG393-2003: X and gamma radiation dose equivalent (rate) meters and monitors used in radiation protection[S]. 2003.