Efficiency calibration of HPGe detector used in water γ radioactivity measurement
To obtain the activity of nuclide in the radioactive measurement of water
it is necessary to know the detection efficiency of nuclide which may be influenced by many factors.
Purpose
2
It is of great practical significance to carry out the efficiency scale of high purity germanium detection device. The high purity germanium detection device selected is based on a real-time on-line monitoring system for the radioactivity of water bodies.
Methods
2
The water samples in the monitored waters were sampled into the low-lead lead rooms
and the radionuclides in water were detected by high purity germanium detectors. The
40
K liquid source with a specific activity was prepared using KCl. The γ spectrum was measured and the detection efficiency of
40
K was calculated. At the same time
the whole water body monitoring system was modeled and calculated according to the specific parameters given by the manufacturers. The energy efficiency of the γ ray in the range of 59.5-2 614.7 keV was simulated and the efficiency curve was obtained.
Results
2
The deviation of detection efficiency of
40
K is found to be 0.018% between the simulation and experimental results.
Conclusion
2
The results verify the feasibility and accuracy of Monte Carlo method in simulating the efficiency scale of high purity germanium detection devices.
Binquan ZHANG , Jizeng MA , Jianping CHENG , . Monte Carlo method is used to calculate the efficiency of high purity germanium detectors for low energy photon measurements . Nuclear Electronics and Detection Technology , 2005 . 25 ( 3 ): 274 - 277 . http://kns.cnki.net/KCMS/detail/detail.aspx?filename=here200503013&dbname=CJFD&dbcode=CJFQ .
WU Xiangyu. Monte Carlo simulation of response function and detection efficiency of γ ray detectors with different sizes[D]. Chengdu:Chengdu University of Technology, 2009.
D Karamanis , V Lacoste , S Andriamonje , . Experimental and simulated efficiency of a HPGe detector with point-like and extended sources . Nuclear Instruments and Methods in Physics Research , 2002 . A487 477 - 487 . https://www.sciencedirect.com/science/article/pii/S0168900202003935 .
A Saraiva , C Oliveira , M Reis , . Study of the response of an ORTEC GMX45 HPGe detector with a multiradio nuclide volume source using Monte Carlo simulations . Applied Radiation and Isotopes , 2016 . 113 47 - 52 . DOI: 10.1016/j.apradiso.2016.04.016 http://doi.org/10.1016/j.apradiso.2016.04.016 .