Zhenhui DAI, Shiwu ZHAO, Geng YANG, et al. Monte Carlo simulation and validation of the enhanced dynamic wedge of Varian accelerator[J]. Nuclear techniques, 2017, 40(12): 120302
DOI:
Zhenhui DAI, Shiwu ZHAO, Geng YANG, et al. Monte Carlo simulation and validation of the enhanced dynamic wedge of Varian accelerator[J]. Nuclear techniques, 2017, 40(12): 120302 DOI: 10.11889/j.0253-3219.2017.hjs.40.120302.
Monte Carlo simulation and validation of the enhanced dynamic wedge of Varian accelerator
Enhanced dynamic wedge (EDW) is commonly used to achieve dose uniformity in the target volume in radiotherapy. However
EDW must be verified before implementation in the treatment planning system.
Purpose
2
This paper is to simulate the EDW of Varian linear accelerator by Monte Carlo method
and to construct a new tool for quality assurance of EDW.
Methods
2
EDW is simulated with the DYNJAWS module of BEAMnrc to build the source model. Depth dose and off-axis dose are calculated by use of DOSXYZnrc in a phantom to generate the EDW factor and profiles of EDW at different angles. The model is validated based on the comparison of Monte Carlo simulation and real measurement results.
Results
2
The simulation results are in good agreement with measurement results
and the deviation is less than 2%.
Conclusion
2
It demonstrates that the EDW model based on Monte Carlo simulation is effective
which would achieve the requirement of quality assurance of EDW.
关键词
Keywords
references
P W Koken , S Heukelom , J P Cuijpers , . On the practice of the clinical implementation of enhanced dynamic wedges . Medical Dosimetry , 2003 . 28 ( 1 ): 13 - 19 . DOI: 10.1016/S0958-3947(02)00140-1 http://doi.org/10.1016/S0958-3947(02)00140-1 .
P Alaei , P D Higgins . Performance evaluation and quality assurance of Varian enhanced dynamic wedges . Journal of Applied Clinical Medical Physics , 2006 . 7 ( 1 ): 14 - 20 . DOI: 10.1120/jacmp.2027.25367 http://doi.org/10.1120/jacmp.2027.25367 .
Lang YU , Bo YANG , Nan LI , . Dosimetry verification of virtual wedge using a two-dimensional ionization chamber array . Chinese Journal of Radiation Oncology , 2015 . 24 ( 3 ): 340 - 343 . DOI: 10.3760/cma.j.issn.1004-4221.2015.03.027 http://doi.org/10.3760/cma.j.issn.1004-4221.2015.03.027 .
D W O Rogers , B A Faddegon , G X Ding , . BEAM:a Monte Carlo code to simulate radiotherapy treatment units . Medical Physics , 1995 . 22 ( 5 ): 503 - 524 . DOI: 10.1118/1.597552 http://doi.org/10.1118/1.597552 .
Qili XIONG , Yong SHI , Gang XU , . Dosimetry characteristics of multi-leaf collimator field for TrueBeam . Nuclear Techniques , 2015 . 38 ( 10 ): 100203 DOI: 10.11889/j.0253-3219.2015.hjs.38.100203 http://doi.org/10.11889/j.0253-3219.2015.hjs.38.100203 http://www.j.sinap.ac.cn/hejishu/CN/abstract/abstract493.shtml .
M B Kakakhel , E S Baveas , A L Fielding , . Validation and automation of the DYNJAWS component module of the BEAMnrc Monte Carlo code . Australasian Physical and Engineering Sciences in Medicine , 2011 . 34 ( 1 ): 83 - 90 . DOI: 10.1007/s13246-011-0060-x http://doi.org/10.1007/s13246-011-0060-x .
F Verhaegen , H H Liu . Incorporating dynamic collimator motion in Monte Carlo simulations:an application in modelling a dynamic wedge . Physics in Medicine and Biology , 2001 . 46 ( 2 ): 287 - 296 . DOI: 10.1088/0031-9155/46/2/302 http://doi.org/10.1088/0031-9155/46/2/302 .
Liang CHEN , Yumin LI , Hong XU . Characteristics of wedge factor of an enhanced dynamic wedge on Varian 600C accelerator . Chinese Journal of Radiation Oncology , 2004 . 13 ( 1 ): 53 - 57 . DOI: 10.3760/j.issn:1004-4221.2004.01.013 http://doi.org/10.3760/j.issn:1004-4221.2004.01.013 .
J P Gibbons . Calculation of enhanced dynamic wedge factors for symmetric and asymmetric photon fields . Medical Physics , 1998 . 25 ( 8 ): 1411 - 1418 . DOI: 10.1118/1.598313 http://doi.org/10.1118/1.598313 .
S A Oh , S K Kim , M K Kang , . Dosimetric verification of enhanced dynamic wedges by a 2D ion chamber array . Journal of the Korean Physical Society , 2013 . 63 ( 11 ): 2215 - 2219 . DOI: 10.1118/1.4814791 http://doi.org/10.1118/1.4814791 .
A W Beavis , S J Weston , V J Whitton . Implementation of the Varian EDW into a commercial RTP system . Physics in Medicine and Biology , 1996 . 41 ( 9 ): 1691 - 1704 . DOI: 10.1088/0031-9155/41/9/009 http://doi.org/10.1088/0031-9155/41/9/009 .
M Ahmad , J Deng , M W Lund , . Clinical implementation of enhanced dynamic wedges into the pinnacle treatment planning system:Monte Carlo validation and patient-specific QA . Physics in Medicine and Biology , 2009 . 54 ( 2 ): 447 - 465 . DOI: 10.1088/0031-9155/54/2/018 http://doi.org/10.1088/0031-9155/54/2/018 .
J H Moeller , D D Leavitt , E Klein . The quality assurance of enhanced dynamic wedges . Medical Dosimetry , 1997 . 22 ( 3 ): 241 - 246 . DOI: 10.1016/S0958-3947(97)00026-5 http://doi.org/10.1016/S0958-3947(97)00026-5 .