Zhenchao ZHANG, Guoping ZUO, Junwen TAN, et al. Locating radioactive source based on three coupled-crystals directional detector of γ rays[J]. Nuclear techniques, 2017, 40(10): 100402
DOI:
Zhenchao ZHANG, Guoping ZUO, Junwen TAN, et al. Locating radioactive source based on three coupled-crystals directional detector of γ rays[J]. Nuclear techniques, 2017, 40(10): 100402 DOI: 10.11889/j.0253-3219.2017.hjs.40.100402.
Locating radioactive source based on three coupled-crystals directional detector of γ rays
,BGO)三种晶体与铅耦合组成的γ射线方向探测器,并采用基于蒙特卡罗方法的通用软件包MCNP(Monte Carlo N Particle Transport Code)研究了铅晶比例、射线能量、剂量率等因素对探测器角度分辨率的影响。结果表明,对于
137
Cs源,在空气吸收剂量率≥ 0.331μGy·h
-1
处,定位角度偏差≤ 0.99°;对于
60
Co源,在空气吸收剂量率0.586 μGy·h
-1
处,测量的平均角度偏差为0.46°;对于水平距离7 m、高度4 m的3.7×10
7
Bq
137
Cs源,相对定位偏差约为5%。
Abstract
Background
2
With the increase application of radioactive sources
the possibility of radiation accidents is raised as well in recent years.
Purpose
2
This study aims to fastly locate lost radioactive sources using a directional detector of γ rays.
Methods
2
Three different coupled scintillators
i.e.
NaI
CsI
Bi
4
Ge
3
O
12
(BGO)
together with center lead cylinder were employed to build the γ ray directional detection system. Localizing spatial radioactive source and the influences of angular resolution caused by lead-crystal ratio
rays energy and source dose rate were simulated by using MCNP (Monte Carlo N particle transport code).
Results
2
The angle deviation is less than 0.99°
provided that the air absorbed dose rate of
137
Cs source is not lower than 0.331 μGy·h
-1
. The average angle measurement deviation is 0.46° when the air absorbed dose rate of
60
Co source is 0.586 μGy·h
-1
. The positioning deviation of 3.7×10
7
Bq
137
Cs source at a distance of 7 m and height of 4 m is about 5%.
Conclusion
2
The three-coupled-crystals directional detector of γ rays can be used to identify the direction of incident γ rays and locate the spatial position of unknown radioactive source with reasonable radiant intensity.
ZHOU Qifu. The situation and countermeasure of radiation safety and security in China[C]. The Council Collected Papers of Nuclear Technology and Industry Application in 2007 Annual Conference, Hubei, 2007:2-9.
Shengen FAN . General situation of radiation accidents in China and analysis of their causes . Radiation Protection , 2002 . 22 ( 5 ): 227 - 281 . DOI: 10.3321/j.issn:1000-8187.2002.05.005 http://doi.org/10.3321/j.issn:1000-8187.2002.05.005 .
M J Willis , S E Skutnik , H L Hall . Detection and positioning of radioactive sources using a four-detector response algorithm . Nuclear Instruments & Methods in Physics Research A , 2014 . 767 ( 11 ): 445 - 452 . DOI: 10.1016/j.nima.2014.08.033 http://doi.org/10.1016/j.nima.2014.08.033 .
R A Powsner , E R Powsner . Essential nuclear medicine physics , 2nd ed : UK Blackwell Publishing , 2008 .
Junwen TAN , Guoping ZUO , Jianliang ZHOU , . The simulation design of a positioning collimator to locate the gamma radiation source with NaI detector . Nuclear Electronics & Detection Technology , 2015 . 35 ( 4 ): 354 - 360 . DOI: 10.3969/j.issn.0258-0934.2015.04.010 http://doi.org/10.3969/j.issn.0258-0934.2015.04.010 .
K A A Gamage , M J Joyce , G C Taylor . Investigation of three-dimensional localization of radioactive sources using a fast organic liquid scintillator detector . Nuclear Instruments & Methods in Physics Research A , 2013 . 707 ( 7 ): 123 - 126 . DOI: 10.1016/j.nima.2012.12.119 http://doi.org/10.1016/j.nima.2012.12.119 .
Y Shirakawa . Development of a direction finding gamma-ray detector . Nuclear Instruments & Methods in Physics Research A , 2007 . 263 ( 1 ): 58 - 62 . DOI: 10.1016/j.nimb.2007.04.056 http://doi.org/10.1016/j.nimb.2007.04.056 .
G L Randall , E Iglesis , H F Wong , . A method of providing directionality for ionising radiation detectors-RadICAL . Journal of Instrumentation , 2014 . 9 ( 10 ): p10011 DOI: 10.1088/1748-0221/9/10/p10011 http://doi.org/10.1088/1748-0221/9/10/p10011 .
Guoping ZUO , Junwen TAN , Jianliang ZHOU , . Study of radioactive source localization based on triangular cylindrical lead shield NaI detector . Atomic Energy Science and Technology , 2017 . 51 ( 3 ): 515 - 521 . DOI: 10.7538/yzk.2017.51.03.0515 http://doi.org/10.7538/yzk.2017.51.03.0515 .
K Fujimoto . A simple gamma ray direction finder . Health Physics , 2006 . 91 ( 1 ): 29 - 35 . DOI: 10.1097/01.HP.0000196113.49929.be http://doi.org/10.1097/01.HP.0000196113.49929.be .