Kai FENG, Yue XU, Hong BAO, et al. Effect of crystal surface characteristics on the performance of PET detectors[J]. Nuclear techniques, 2017, 40(6): 060402
DOI:
Kai FENG, Yue XU, Hong BAO, et al. Effect of crystal surface characteristics on the performance of PET detectors[J]. Nuclear techniques, 2017, 40(6): 060402 DOI: 10.11889/j.0253-3219.2017.hjs.40.060402.
Effect of crystal surface characteristics on the performance of PET detectors
Positron emission tomography (PET) is the most sensitive and quantitative functional and molecular imaging technique. It was more and more popularly used in modern biomedical research such as animal model of human disease
new drug development and evaluation of new therapies.
Purpose
2
Improving the detector property is the key to increase the performance of the PET scanners. PET detector normally consists of a segmented scintillator array and a photodetector.
Methods
2
In this work the performance of PET detectors consisting of (Lu
Y)
2
SiO
5
(LYSO) arrays with polished and unpolished crystal surface was measured by using both dual-ended and single-ended readout with position sensitive photomultiplier tube.
Results
2
The experimental results showed that both LYSO arrays provided similar flood histogram and energy resolution for dual-ended readout detector
but the unpolished LYSO array provided much better resolution of interaction depth and should be used for dual-ended readout PET detectors.
Conclusion
2
Polished LYSO array provided better flood histogram and energy resolution
and should be used for single-ended readout PET detector.
关键词
Keywords
references
S R Cherry , Y Shao , S Siegel , . MicroPET:a high resolution PET scanner for imaging small animals . IEEE Transactions on Nuclear Science , 1996 . 44 ( 3 ): 1161 - 1166 . https://www.researchgate.net/publication/3135578_MicroPET_A_high_resolution_PET_scanner_for_imaging_small_animals .
Zhonghua KUANG , Cheng LI , Lanjun LI , . Research progress on PET with high resolution and high sensitivity . Nuclear Physics Review , 2016 . 33 ( 3 ): 336 - 344 . http://www.cnki.com.cn/Article/CJFDTOTAL-HWDT201603016.htm .
W W Moses . Fundamental limits of spatial resolution in PET . Nuclear Instruments and Methods in Physics Research A , 2011 . 648 ( Suppl 1 ): S236 - S240 . http://adsabs.harvard.edu/abs/2011NIMPA.648S.236M .
Hong YIN , Yang XU , Dehui LI , . Study on LYSO scintillation crystal array for small animal PET applications . Piezoelectrics and Acoustooptics , 2014 . ( 3 ): 406 - 408, 411 . DOI: 10.11977/j.issn.1004-2474.2014.03.024 http://doi.org/10.11977/j.issn.1004-2474.2014.03.024 .
A J Gonzalez , A Aguilar , P Conde , . A PET design based on SiPM and monolithic LYSO crystals:performance evaluation . IEEE Transactions on Nuclear Science , 2016 . 63 ( 5 ): 2471 - 2477 . DOI: 10.1109/TNS.2016.2522179 http://doi.org/10.1109/TNS.2016.2522179 .
Y F Yang , S S James , Y B Wu , . Tapered LSO arrays for small animal PET . Physics in Medicine and Biology , 2011 . 56 ( 1 ): 139 - 153 . DOI: 10.1088/0031-9155/56/1/009 http://doi.org/10.1088/0031-9155/56/1/009 .
CHEN Ze, HU Zhengguo, CHEN Jinda, et al . Simulation study on the properties of LYSO crystal strip used in the TOF-PET prototype[C]. Harbin:Progress Report on China Nuclear Science and Technology, 2013.
Y C Tai , A F Chatziioannou , Y F Yang , . MicroPET Ⅱ:design, development and initial performance of an improved microPET scanner for small-animal imaging . Physics in Medicine and Biology , 2003 . 48 ( 11 ): 1519 - 1537 . DOI: 10.1088/0031-9155/48/11/303 http://doi.org/10.1088/0031-9155/48/11/303 .
Y F Yang , Y B Wu , J Y Qi , . A prototype PET scanner with DOI-encoding detectors . Journal of Nuclear Medicine , 2008 . 49 ( 7 ): 1132 - 1140 . DOI: 10.2967/jnumed.107.049791 http://doi.org/10.2967/jnumed.107.049791 .
Y F Yang , P A Dokhale , R W Silverman , . Depth of interaction resolution measurements for a high resolution PET detector using position sensitive avalanche photodiodes . Physics in Medicine and Biology , 2006 . 51 ( 9 ): 2131 - 2142 . DOI: 10.1088/0031-9155/51/9/001 http://doi.org/10.1088/0031-9155/51/9/001 .
Cheng LI , Zhonghua KUANG , Junwei DU , . High spatial resolution and high sensitivity three dimensional PET detector development with position-sensitive avalanche photodiode and silicon photomultiplier readout . Journal of Integration Technology , 2016 . 5 ( 4 ): 1 - 13 . http://www.cnki.com.cn/Article/CJFDTOTAL-JCJI201604001.htm .