WANG Yanbing, LIU Xialing, QIN Bin, et al. Integral magnetic field measurement of THz-FEL oscillator undulator based on stretch-wire method[J]. Nuclear techniques, 2015, 38(7): 070103
DOI:
WANG Yanbing, LIU Xialing, QIN Bin, et al. Integral magnetic field measurement of THz-FEL oscillator undulator based on stretch-wire method[J]. Nuclear techniques, 2015, 38(7): 070103 DOI: 10.11889/j.0253-3219.2015.hjs.38.070103.
Integral magnetic field measurement of THz-FEL oscillator undulator based on stretch-wire method
Background: In collaboration with the National Synchrotron Radiation Laboratory
Huazhong University of Science and Technology is developing an oscillator-based THz free-electron laser (HUST THz-FEL). Purpose: In order to measure the magnetic field integrals of the undualtor accurately and quickly
a measurement system based on the stretch-wire method (SWM) was established. Methods: In the system
an eight and a half digital multimeter
Agilent 3458
was adopted to acquire the micro-volt voltage signal
and a Kohzu high precision positioning platform was employed to realize the high synchronized wire movement. The software of this system was implemented by LabVIEW programming Results: The field measurement system was tested in a prototype hybrid permanent undulator with 7 periods
which shows consistent results with the Hall probe mapping. The repeatability precision for first integral and second integral are better than 2.5?10?6 T?m and 2.5?10?6 T?m2 respectively. Conclusion: The system met the design requirements and will be applied to the compact THz FEL oscillator located in Huazhong University of Science and Technology.Cited