Influencing factors of dose equivalence for X and γ rays with different energy based on Monte Carlo
Background: The accuracy of dosimeter measurement of X and γ rays needs to be resolved. Purpose: The aim is to study the correction term of the equivalent process of low-energy X-ray and the natural radioactive source. Methods: Instead of the standard sources
X-ray machine was adopted on the dose instrument calibration. The influence factors of the equivalence between low-energy X-ray and high-energy X or γ rays were simulated using Monte Carlo (MCNP) software. Results: The influences of distance
space scattering
response of detector on dose equivalence were obtained. The simulation results were also analyzed. Conclusion: The method can be used in dose equivalent correction for low-energy X-ray
high-energy X or γ rays