LI Qi WANG Shilian FAN Yuanqing CHANG Yinzhong LIU Shujiang SHI Jianfang ZHANG Xinjun CHEN Zhanying JIA Huaimao ZHAO Yungang. Study on split method of aerosol sample from IMS radionuclide station[J]. Nuclear techniques, 2013, 36(1): 010204
LI Qi WANG Shilian FAN Yuanqing CHANG Yinzhong LIU Shujiang SHI Jianfang ZHANG Xinjun CHEN Zhanying JIA Huaimao ZHAO Yungang. Study on split method of aerosol sample from IMS radionuclide station[J]. Nuclear techniques, 2013, 36(1): 010204 DOI: 10.11889/j.0253-3219.2013.hjs.36.010204.
Study on split method of aerosol sample from IMS radionuclide station
Level 5 samples from radionuclide stations of the Comprehensive Nuclear Test Ban Treaty(CTBT) should be split and sent to two or more independent laboratories for measurement and analysis.The study is focused on the sample splitting method and the uniformity testing of split samples.Two sample splitting methods
filter cutting and sample compressing split
are introduced in this article.Uncompressed and compressed filter samples are split by using two methods.Then the split samples are measured by HPGe γ spectrometer.The results show that the split samples are homogeneous.This work establishes the split and test foundation of the radionuclide station and laboratory.