sing Sophy DSX SPECT and 32P or 90y as the radioactive sources
the optimal bremsstrahlung imaging condition with different collimators
various widths of energy window
and different distance etc was discussed.The results showed that the bremsstrahlung spectrum was continious
the satisfactory images can be obtained with high-energy collimator and 80keV-60% energy window scale or empirical-window;when the distance between collimator and source was short.Preliminary clinical result was also satisfactory.Thereby
it was suggested that SPECT can be applied to bressmtrahlung imaging.