浏览全部资源
扫码关注微信
1. 北京协和医院
2. 北京协和医院,北京,100730
Published:01 November 1993
移动端阅览
[1]赵周社,周前.SPECT图像三种散射校正方法的比较[J].核技术,1993(11):674-678.
Zhao Zhoushe Zhou Qian. Comparison of three scatter correction techniques in SPECT imaging[J]. Nuclear techniques, 1993, (11): 674-678.
对SPECT图像的单能窗、双能窗及多能窗加权采集三种散射校正方法进行比较研究。模型实验及临床试验结果均为单能窗(对称窗)采集时对比度及FSD%(标准差分数百分值)最小
多能窗加权采集时对比度及FSD%均最大
双能窗采集时对比度和FSD%介于二者之间。虽然多能窗加权采集法从对比度和FSD%综合考虑明显优于单能窗及双能窗采集法
但是该方法计算过程复杂、占用内存大。双能窗采集法相对简单、结果可靠
更适合于临床常规应用。
Three scatter correction techniques including (Ⅰ) single energy window
(Ⅱ) dual energy window
and (Ⅲ) multi- energy window weighted acquisition were compared in this study. Results indicated that the contrast and percentage fraction standard deviation(FSD%) were smallest in Ⅰ and largest in Ⅲ
as calculated from phantom test and clinical application. Although the result of Ⅲ is the best
its calculation is very complicated and occupies more disk space. It concludes that Ⅱ is more suitable for routine clinical use
due to its simplicity and accuracy.