Wu Chanqun Jiang yimin Li Yinmin. Establishment of radioimmunoassay for transferrin and its preliminary application[J]. Nuclear techniques, 1993, (11): 653-656.
Wu Chanqun Jiang yimin Li Yinmin. Establishment of radioimmunoassay for transferrin and its preliminary application[J]. Nuclear techniques, 1993, (11): 653-656.DOI:
Establishment of radioimmunoassay for transferrin and its preliminary application
A guantivitive radioimmunoassay of transferrin was established. Its antisera was obtained by immunizing rabbits with purified transferrin and the titer was as high as 1:6×104 with an affinity constant of 1.15×1012 l/mol. Ferrin
lactoferrin and serum aibimin showed no significant interference. The sensitivity was 2.54 ng/tube and the recovery was X =102.91%. The inter-batch CV was 1.8% (n = 14) and intra-batch CV was 9.94% (n=10). The transferrin level (X±SD) of serum
urine and saliva in healthy adults measured by this method were 3.03±0.66 g/l (n = 31)
198±98 μg/l (n=55) and 557.89±206.32 μg/l (n=12) respectively. Serum transferrin in iron deficiency anemia were 6.46±4.15 g/l (n= 134). 29 out of 35 cases of patients with abnormai. renal function had their urine transferrin higher than normal.