浏览全部资源
扫码关注微信
1. 中山大学
2. 中山大学,广州,510275
Published:01 January 1992
移动端阅览
[1]罗达玲,黄乃明,张纯祥.LiF热释光剂量计的剂量响应的研究[J].核技术,1992(01):51-58.
Luo Daling Huang Naiming Zhang Chunxiang. Dose response of thermoluminescent lithium fluoride to gamma radiation[J]. Nuclear techniques, 1992, (1): 51-58.
我们运用国产FJ377热释光读出装置与APPLE-Ⅱ微机组成联机系统
配合微机解谱技术测得LiF(Mg
Cu
P)粉末样品3个发光峰所对应的γ辐射剂量响应均为亚线性。本文以统计原理为基础
提出了用一次击中和二次击中相叠加的剂量响应函数描述LiF热释光剂量计的剂量响应
并将LiF(Mg、Ti)的超线性和LiF(Mg
Cu
P)的亚线性统一用一个公式表示
给出了剂量响应的线性、亚线性和超线性与特征剂量D
0
和R的关系。模型计算与实验符合很好。
Dose responses of thermoluminescent LiF(Mg
Cu
P) to gamma rays radiation have been investigated with a FJ-377 TLD reader which is interfaced to an Apple Ⅱ microcomputer. Sublinear responses of glow peak 1
peak2 and peak3 of the TLD have been obtained. Based on statistical distribution and hit concepts in microdosimetry
a function of TL response to gamma rays was proposed. The function implying a combination of 1-hit and 2-hit responses is a very good decription of TL response of linerarity
sublinearity and supraliearity. Parameters used in this model consist of characteristic dose Do and faction R for 1-hit detectors. Experimental results are in good agreement with the model calculation.