WANG Liyan LI Min WU Zhangwen GOU Chengjun HOU Qing. M-C simulation of photon fluence distribution in IMRT by electromagnetic scanning low-energy E-beams[J]. Nuclear techniques, 2010, 33(10): 760-764.
WANG Liyan LI Min WU Zhangwen GOU Chengjun HOU Qing. M-C simulation of photon fluence distribution in IMRT by electromagnetic scanning low-energy E-beams[J]. Nuclear techniques, 2010, 33(10): 760-764.DOI:
M-C simulation of photon fluence distribution in IMRT by electromagnetic scanning low-energy E-beams
an approach to generate uniform photon beams is to use electron pencil beams to bombard a set of points that are discretely arranged on the target. It was considered that the E-beam energy in this technique should be greatly higher than that those from accelerators commonly available for radiation therapy. In this study
Monte Carlo simulation was used to explore the possibility of generating the needed photon fluence distribution by using a common medical accelerator. The fluence distributions produced by electrons pencil beam in different energies were compared. Optimal target materials
and the geometry
were determined. The photon fluence distributions of wide fields at various target point intervals were simulated for suitable density of the target points. Two fluence distributions in common request by conventional IMRT were taken as examples to show that the desired fluence distributions could be achieved by optimizing the intensity of electron beam on target points.
Optimization study on the size of medical radioactive waste liquid sampling container
Optimization of segment weight using simulated dynamics algorithm for beamlet-based IMRT
Related Author
LU Xiaojun
LIU Jiayu
HU Chongqing
HAN Gang
SONG Jiaban
HE Linfeng
CHANG Jinxue
CHEN Bingzhou1
Related Institution
Shanghai Institute of Measurement and Testing Technology
Key Lab for Radiation Physics and Technology of Ministry of Education , Institute of Nuclear Science and Technology, Sichuan University,Chengdu 610064) 2