FANG Chunhui1, 3 FANG Yan1, 3 JIA Quanjie2, et al. A method for X-ray measurement of liquid structure by reflection of synchrotron radiation[J]. Nuclear techniques, 2007, (7): 560-564.
FANG Chunhui1, 3 FANG Yan1, 3 JIA Quanjie2, et al. A method for X-ray measurement of liquid structure by reflection of synchrotron radiation[J]. Nuclear techniques, 2007, (7): 560-564.DOI:
A method for X-ray measurement of liquid structure by reflection of synchrotron radiation
An experimental method is reported in this paper for X-ray measuringement of liquid structure by X-ray reflection of from flat specimen cell with a the five-circle goniometer in of the 4W1C beamline of Beijing Synchro-tron Radiation Facility(BSRF I) was reported in the paper.Taking the scattering measurement of an aqueous solution of sodium sulfate for as an example
construction of liquid cell
scattering geometrical arrangement
sample alignment
and scanning mode were briefly introduced.Corrections of the measured scattering intensity for such as net absorp-tion of beryllium window
air scattering
synchrotron radiation polarization
sample absorption and geometrical factor
etc.were described in detail.Calculation procedures of liquid structural function and radial distribution function
in-cluding removal of Compton scattering and spurious peak
were demonstrated.The result shows that high quality X-ray scattering data can be obtained by the X-ray reflection method.