WANG Xuetao1, 2 CHEN Shaowen3 DAI Zhenhui2 ZHU Lin2 LIU Xiaowei1 1. Angular dependence study on dose distribution of MatriXX 2-D chamber array[J]. Nuclear techniques, 2012, 35(2): 126-130.
WANG Xuetao1, 2 CHEN Shaowen3 DAI Zhenhui2 ZHU Lin2 LIU Xiaowei1 1. Angular dependence study on dose distribution of MatriXX 2-D chamber array[J]. Nuclear techniques, 2012, 35(2): 126-130.DOI:
Angular dependence study on dose distribution of MatriXX 2-D chamber array
Monte Carlo计算和MatriXX测量的曲线及平面中心点剂量均存在明显差异。研究结果为多机架角度照射合成验证调强放疗计划的剂量分布提供依据。
Abstract
With the gantry angles of 0° 80° and 280° 360°(0°)
the off axis distance(OAR) profiles curves of 6 MV X-ray were obtained by Monte Carlo simulation
3-D water phantom measurement and MatriXX measurement.Comparing the three types of curves
we analyzed the angular dependence of MatriXX 2-D chamber array.The results revealed that the three types of curves agreed well with the gantry angles of 0° 60° and 300° 360°(0°).Curves and centre point dose on plane from Monte Carlo simulation differed obviously from those of MatriXX measurement in gantry angles of 70° 90° and 270° 290°.It is possible to verify the dose distribution of intensity modulated radiation therapy(IMRT) plan using the Multi-Gantry-Angle Composite(MGAC).