137Cs and 60Co in mixed point sources places at different radius positions in empty sample boxes of different heights were obtained.The peak efficiency function parameters of the point sources were determined by least-squares fitting method.The 81.0
160.6
163.9
233.2
249.8
526.6 keV γ-ray detection efficiencies of Φ63.50 mm×16.66 mm xenon samples were obtained by numerical integration.This method was used to calculate activity of collected xenon samples.The results were
compared with those by Monte Carlo simulation and ?-? energy coincidence measurement