This paper describes the use of TLD method to measure exposure dose rate in
60
Co garden. As compared with chemical dosimetry
the relative deviation of measurement would not exceed 10%. The good repeatability has been proved by comparing the measurements of different times and distances with correction of decay law and 1/1
2
law
Besides
the distribution of exposure dose rate was influenced by scattering of the living plants.
关键词
Keywords
references
N. W. Holm and R. J. Berry , Manual on radiation dosimetry , New York , Deket , 1970 .
S. C. Bushong et al. , Heal. Phys. , 26 . 358 ( 1974 ).