A cryogenic assembly of Mössbauer spectrometer is introduced. Its applicable temperature range is from 77K to 300K. The temperature controlling precision is ±0.1K and the temperature measuring precision is ±0.6K. The capacity of Dewar container is two liters. The consumptions of liquid N
2
are 0.05 1/h and 0.14 1/h at 77K and 273K respectively. This cryogenic assembly is sufficiant to meet the demands of Mössbauer spectrum measurement.
关键词
Keywords
references
贡泽尔 编, 徐英庭 等 译 , 穆斯堡尔谱学 , 科学出版社 , 1979 年,第 36 页。
T. C. Gibb , Principles of Mössbauer spectroscopy , John Wiley and Sons Inc. , New York , 1976 , p. 135 .