The ampoule-type Fricke dosimeter as an standard of absorbed dose
|更新时间:2022-08-16
|
The ampoule-type Fricke dosimeter as an standard of absorbed dose
Nuclear TechniquesVol. 13, Issue 3, (1990)
作者机构:
江苏省农科院原子能所,南京辐照中心
作者简介:
基金信息:
DOI:
CLC:
Received:28 March 1988,
Published:01 March 1990
稿件说明:
移动端阅览
Naining Li, Daochuan Yin, Yunsheng Zhou, et al. The ampoule-type Fricke dosimeter as an standard of absorbed dose[J]. Nuclear techniques, 1990, 13(3): 176-181.
DOI:
Naining Li, Daochuan Yin, Yunsheng Zhou, et al. The ampoule-type Fricke dosimeter as an standard of absorbed dose[J]. Nuclear techniques, 1990, 13(3): 176-181.DOI:
The ampoule-type Fricke dosimeter as an standard of absorbed dose
The ampoule-type Fricke dosimeter and factors that affect its uses were discussed in this paper. The molar extinction coefficient for Fe
3+
at 302 nm was determined to be 2166.7 1/mol·cm (25℃) and the conversion factor for absorbence-absorbed lose was obtained to be 280.7Gy/∆Abs. The technique of pretreatment
sealing of glass ampoule
and prepartion of Fricke dosimeter were standardized
so that the precision of measured valuse has been effectively controlled within ±1.0% (relative standard deviation of an average value at 95% confidence level)