Studies on dichromate solution/Ag+ dosimeters as 60Co absorption dose transfer standard in the 10—40 kGy range
|更新时间:2022-08-16
|
Studies on dichromate solution/Ag+ dosimeters as 60Co absorption dose transfer standard in the 10—40 kGy range
Nuclear TechniquesVol. 13, Issue 1, (1990)
作者机构:
广州分析测试中心
作者简介:
基金信息:
DOI:
CLC:
Received:25 January 1988,
Published:01 January 1990
稿件说明:
移动端阅览
Zhutang Wen, Riyou Luo, Lin Gao, et al. Studies on dichromate solution/Ag+ dosimeters as 60Co absorption dose transfer standard in the 10—40 kGy range[J]. Nuclear techniques, 1990, 13(1): 29-33.
DOI:
Zhutang Wen, Riyou Luo, Lin Gao, et al. Studies on dichromate solution/Ag+ dosimeters as 60Co absorption dose transfer standard in the 10—40 kGy range[J]. Nuclear techniques, 1990, 13(1): 29-33.DOI:
Studies on dichromate solution/Ag+ dosimeters as 60Co absorption dose transfer standard in the 10—40 kGy range
dosimeters in the 10—40 kGy range the response to dose is in good linearity
and the effects of dose rate and temperature during irradiation on absorption dose response could be negligible. Furthermore
the dosimeter solutions have good storage stability before and after irradiation. So long as the preparation of dosimeters and the measuring procedure are well controlled
the precision of the system can reach to about 1%
and total uncertainty to ±3.5%. The system can be used as transfer standard dosimeters for high-dose measurement.
关键词
Keywords
references
R. W. Matthews , Int. J. Appl. Radiat. Isotope , 32 ( 1981 ), 861 .
P. H. G. Sharpe et al . , Int. J. Appl. Radiat. Isotope , 36 ( 1985 ), 647 .
J. Thomassen , Internationl symposium on high-dose dosimetry, IAEA-SM-272/35 , IAEA , ( 1984 ).