construction and characteristics of the Bragg curve spectroscopy detector (BCS detector) are described in this paper.
Our BCS detector with its field parallel to the particle trajectories was tested by 8.78MeV and 6.02MeV α particles from a ThC-ThC' source. The energy resolutions of 1.5% and 2.6% respectively for two group of α particles and the charge resolution of 2.7% were obtained. Further test experiments with heavy ions will be arranged.
关键词
Keywords
references
С. R. Gruhn et al ., Nucl. Instr. Meth. 196 , 33 ( 1982 ).
Ch. Schiessl et al ., Nucl Instr. Meth ., 192 , 291 ( 1982 ).
J. M. Asselineau et al ., Nucl. Instr. Meth ., 204 , 109 ( 1982 ).
H. Ikezoe et al ., Nucl. Instr. Meth ., 196 , 215 ( 1982 ).
R. J. Mcdonald et al ., Nucl. Instr. Meth ., 219 , 508 ( 1984 ).
M. Arialdo et al ., Nucl. Instr. Meth ., 225 , 57 ( 1984 ).
L. Northcliffe et al ., Nucl. Data Tables , A7 ( 1970 ).
O. Bunemann et al ., Can. J. Res ., 27A , 191 ( 1949 ).
H. W. Fulbright , Nucl. Instr, Meth ., 162 , 21 ( 1979 ).